Pagina 1 - Laboratory pH Meter

Labor-pH-MeterLaboratory pH MeterCG 842ba12242de_2 09/99Gebrauchsanleitung / Operating manualBedienungsanleitung Seite 1Operating manual Page 45

Pagina 2

Overview531.5 Technical data Measuring accuracy(± 1 digit)Given as measuring uncertainty at a confidence level of 95%. All accuracies additionally de

Pagina 3

Overview54 1 Keypad Foil keypad (Polyester)Length [mm] 230Width [mm] 210Height [mm] 70Weight [kg] Approx. 0.850Material ABSInterference emission (

Pagina 4

Overview55Batteries 4 x 1.5 V AA type alkaline manga-nese batteriesRuntime Approx. 3000 operating hoursMains power supply (option)Connection max. over

Pagina 5 - List of contents

Overview56Electrodes Socket according to DIN 19 262 or BNC4 mm socket for reference elec-trodeTemperature probe 4 mm socket fortemperature probePlug-i

Pagina 6 - 1Overview

Safety572 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation and main-tenance of the pH

Pagina 7 - 1.1 Keyboard

Safety582.1 Authorized useThis instrument is authorized exclusively for pH and Redox measurements in the laboratory. The technical specifications as g

Pagina 8 - 1.3 Sockets

Safety59Safe operation If safe operation is no longer possible, the instrument must be taken out of service and secured against inadvertent op-eration

Pagina 10 - 1.5 Technical data

Commissioning613 CommissioningPerform the following activities for the initial commissioning:l Set the °C or °Fl Set the resolutionl Connect the plug-

Pagina 11 - Overview

Commissioning62Connecting theplug-in powersupply (optional)The measuring instrument works battery-powered. It can, however, also be supplied by the pl

Pagina 12

Laboratory pH MeterCG 842ba12242e_2 09/99Operating Manual 45

Pagina 13

Operation634Operation4.1 Switch on the instrumentNoteThe instrument has an energy saving feature to avoid un-necessary battery depletion. The energy s

Pagina 14 - 2 Safety

Operation644.2 MeasuringPreparatoryactivitiesPerform the following activities when you want to measure:NoteIncorrect calibration of the pH electrode w

Pagina 15 - 2.1 Authorized use

Operation654.2.1 Measuring the pH valueAutoRead AR(Drift control)The AutoRead function (drift control) checks the stability of the measurement signal.

Pagina 16

Operation66NoteThe current AutoRead measurement (with acceptance of the current value) can be terminated at any time by pressing <run/enter>.4.2

Pagina 17

Operation674.3 CalibratingWhy calibrate? pH electrodes age. This changes the asymmetry (zero point) and slope of the pH electrode. As a result, an ine

Pagina 18 - 3 Commissioning

Operation68CalibrationevaluationAfter the calibration, the instrument automatically evaluates the current status. The asymmetry and slope are separate

Pagina 19 - Commissioning

Operation694.3.1 AutoCal TECUse any two of the Technical Buffer Solutions for this proce-dure.NoteSteps 2 and 6 are not required if you use a temperat

Pagina 20 - 4Operation

Operation706 If necessary, set the temperature of the second buffer solution by pressing <▲> <▼>.7 Submerse the electrode in the second bu

Pagina 21 - 4.2 Measuring

Operation714.3.2 AutoCal DINUse two different DIN buffer solutions for this procedure:NoteSteps 2 and 6 are not required if you use a temperature prob

Pagina 22 - AutoRead

Operation725 Thoroughly rinse the electrode with distilled water.6 If necessary, set the temperature of the second buffer solution by pressing <▲&g

Pagina 23 - Operation

Accuracy whengoing to pressAll details contained in this operating manual are valid data at the time of going to press. However, SCHOTT may add detail

Pagina 24 - 4.3 Calibrating

Operation734.3.3 ConCalTwo-point calibrationUse two buffer solutions for this procedure:l pH 7.0 ± 0.5l any other buffer solutionNoteSteps 2 and 9 are

Pagina 25

Operation748 Thoroughly rinse the electrode with distilled water.9 If necessary, set the temperature of the second buffer solution by pressing <▲&g

Pagina 26 - ▲> <▼>

Operation75Single-point calibrationUse a buffer solution in the range pH = 7.0 ± 0.5 for this pro-cedure.NoteOnly the electrode asymmetry is determine

Pagina 27 - ▲> <▼>

Operation764.4 ResetBasic settings The following functions are reset (initialized) to the values they had on delivery:Proceed as follows:Measuring mod

Pagina 28

Maintenance, cleaning, disposal775 Maintenance, cleaning, disposal5.1 MaintenanceThe measuring instrument is almost maintenance-free. The only mainten

Pagina 29

Maintenance, cleaning, disposal785.2 CleaningOccasionally wipe the outside of the measuring instrument with a damp, lint-free cloth. Disinfect the hou

Pagina 30 - ▲> <▼> keys

What to do if...796 What to do if... Cause RemedypH electrode:– Not connected – Connect electrode– Air bubbles in front of the diaphragm– Remove

Pagina 31

What to do if...80 Buffer solutions:– Incorrect buffer solutions – Change calibration procedure– Buffer solutions too old – Only use once.Note the

Pagina 32

What to do if...81 Cause Remedy– pH electrode unsuitable – Use suitable electrode– Temperature difference between buffer and sample too large– Adjust

Pagina 33 - 4.4 Reset

What to do if...82

Pagina 34 - 5.1 Maintenance

List of contents471 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 491.1 Keyboard . . . . . . . . . . . . . . . . . . . . . . .

Pagina 35 - 5.3 Disposal

Lists837ListsThis chapter provides additional information and orientation aids.Abbreviations The list of abbreviations explains abbreviations that app

Pagina 36 - 6 What to do if

Lists84AbbreviationsAR AutoRead (drift control)ARng Automatic range switchingMeasuring instrument measures with high-est resolutionASY AsymmetryAutoCa

Pagina 37 - What to do if

Lists85TP Temperature probeTemperature measurement activeUASY Asymmetry potential°C Temperature unit, °Celsius°F Temperature unit, Fahrenheit

Pagina 38

Lists86GlossaryAsymmetry Zero point of a pH electrode.Resolution Number of decimal places that appear for a measured val-ue.AutoRead Monitors the elec

Pagina 39

Lists 87IndexAasymmetry 67authorized use 58AutoCal DIN 67, 71AutoCal TEC 67, 69AutoRead 65Bbasic settings 76battery compartment 77Buchsenfeld 51Ccalib

Pagina 40

List of contents48

Pagina 41 - HAT TO DO IF ...)

Overview491OverviewThe compact CG 842 precision pH meter lets you perform pH measurements rapidly and reliably. The CG 842 provides the highest degree

Pagina 42 - ASY Asymmetry potential

Overview501.1 KeyboardSwitch measuring instrument ON/OFF<on/off>Select measuring mode<pH>Call up calibration procedure<CAL>Activate/

Pagina 43 - UTOREAD

Overview511.2 Display1.3 Sockets CautionOnly connect probes to the instrument that cannot feed ex-cessive voltages or currents (> SELV and > cir

Pagina 44

Overview521.4 Declaration of conformity29.03.99SCHOTT Geräte GmbHIm Langgewann 5D 65719 Hofheim am TaunusDeutschland, Germany, Allemagne29. März, Marc