OPERATIONS MANUAL
ba76143e01 04/2013
MultiLab 4010-3
pH/ORP/D.O./Cond
DIGITAL METER FOR IDS SENSORS
4010-3
Vista de pagina 0
1 2 3 4 5 6 ... 111 112

Indice de contenidos

Pagina 1 - MultiLab 4010-3

OPERATIONS MANUALba76143e01 04/2013MultiLab 4010-3pH/ORP/D.O./CondDIGITAL METER FOR IDS SENSORS4010-3

Pagina 2 - © 2013 Xylem Inc

Safety MultiLab 4010-310 ba76143e01 04/20132Safety2.1 Safety information2.1.1 Safety information in the operating manualThis operating manual provides

Pagina 3 - MultiLab 4010-3 - Contents

Glossary MultiLab 4010-3100 ba76143e01 04/201316 GlossarypH/ORPConductivityAsymmetry see zero pointElectromotive force ofa combinationelectrodeThe ele

Pagina 4 - 4 ba76143e01 04/2013

MultiLab 4010-3 Glossaryba76143e01 04/2013 101Dissolved oxygenGeneral informationSalinity The absolute salinity SA of seawater corresponds to the rela

Pagina 5

Glossary MultiLab 4010-3102 ba76143e01 04/2013Measured value The measured value is the special value of a measured parameter to be determined. It is g

Pagina 6 - 6 ba76143e01 04/2013

MultiLab 4010-3 Indexba76143e01 04/2013 10317 IndexAAir calibration beaker . . . . . . . . . . . . . . . . . 49AutoRead . . . . . . . . . . . . . .

Pagina 7 - 1 Overview

Index MultiLab 4010-3104 ba76143e01 04/2013Transmitting data . . . . . . . . . . . . . . . . . . . .82Automatic . . . . . . . . . . . . . . . . . .

Pagina 8 - 1.2 Sensors

MultiLab 4010-3 Indexba76143e01 04/2013 105

Pagina 9

Appendix MultiLab 4010-3106 ba76143e01 04/201318 Appendix18.1 Oxygen solubility tableSolubility of oxygen in mg/L in water exposed to water-saturated

Pagina 10 - 2.2 Safe operation

MultiLab 4010-3 Appendixba76143e01 04/2013 10733.0 7.183 6.843 6.518 6.208 5.912 5.63134.0 7.065 6.732 6.415 6.111 5.822 5.54635.0 6.950 6.624 6.314 6

Pagina 11 - ECHNICAL DATA, page 94)

Appendix MultiLab 4010-3108 ba76143e01 04/201318.2 DO% Calibration valuesPressure Altitude Calibration valueInches Hg mm Hg kPa mbar Feet Meters % Sat

Pagina 12 - 3 Commissioning

MultiLab 4010-3 Contact Information109ba76143e01 04/201319 Contact Information19.1 Ordering & Technical SupportWhen placing an order please have t

Pagina 13 - ENERAL DATA

MultiLab 4010-3 Safetyba76143e01 04/2013 11Only the operation and running of the meter according to the instructions and technical specifications give

Pagina 14 - 4 Operation

Contact Information MultiLab 4010-3110ba76143e01 04/2013

Pagina 16 - 16 ba76143e01 04/2013

1) The tissue in plants that brings water upward from the roots;2) a leading global water technology company.We're 12,500 people unified in a com

Pagina 17 - 03.04.2013 08:00

Commissioning MultiLab 4010-312 ba76143e01 04/20133 Commissioning3.1 Scope of delivery MeterMultiLab 4010-3  USB cable (A plug on mini B plug) Powe

Pagina 18

MultiLab 4010-3 Commissioningba76143e01 04/2013 133.3.1 Connecting the power packCAUTIONThe line voltage at the operating site must lie within the inp

Pagina 19 - 4.4 Login with user name

Operation MultiLab 4010-314 ba76143e01 04/20134 Operation4.1 General operating principles4.1.1 KeypadIn this operating manual, keys are indicated by b

Pagina 20 - Change password

MultiLab 4010-3 Operationba76143e01 04/2013 154.1.2 DisplayExample (pH):4.1.3 Status information1 Status information (meter)2 Status information (sens

Pagina 21 - 4.5 Navigation

Operation MultiLab 4010-316 ba76143e01 04/20134.1.4 Instrument connectors1 IDS sensors: (pH, ORP, conductivity, D.O.)a) Channel 1b) Channel 2c) Channe

Pagina 22

MultiLab 4010-3 Operationba76143e01 04/2013 174.1.5 Channel displayThe MultiLab 4010-3 manages the connected sensors and displays which sen-sor is plu

Pagina 23

Operation MultiLab 4010-318 ba76143e01 04/20134.1.7 Display of several sensors in the measuring modeThe measured values of the connected sensors can b

Pagina 24

MultiLab 4010-3 Operationba76143e01 04/2013 194.2 Switching on the meter4.3 Switching off the meter4.4 Login with user nameAfter activation of the use

Pagina 25 - Clock function

2 ba76143e01 04/2013MultiLab 4010-3For the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs, OH

Pagina 26 - 26 ba76143e01 04/2013

Operation MultiLab 4010-320 ba76143e01 04/2013Changing thepasswordIf the administrator has set up the access with password protection:2. Using <&g

Pagina 27 - 5 pH value

MultiLab 4010-3 Operationba76143e01 04/2013 21Forgotten thepassword?Contact the administrator.4.5 Navigation4.5.1 Operating modes4.5.2 Measured value

Pagina 28 - Δ : Better than 0.5 °C

Operation MultiLab 4010-322 ba76143e01 04/2013The selection is made with the <><> keys. The current selection is dis-played with a frame

Pagina 29 - 5.2 pH calibration

MultiLab 4010-3 Operationba76143e01 04/2013 23 MessagesInformation is marked by the [i] symbol. It cannot be selected. Example:4.5.4 Example 1 on nav

Pagina 30 - 30 ba76143e01 04/2013

Operation MultiLab 4010-324 ba76143e01 04/20133. Select the System submenu with <><>. The current selection is displayed with a frame.4.

Pagina 31

MultiLab 4010-3 Operationba76143e01 04/2013 254.5.5 Example 2 on navigation: Setting the date and timeThe meter has a clock with a date function. The

Pagina 32 - 32 ba76143e01 04/2013

Operation MultiLab 4010-326 ba76143e01 04/20134. Change and confirm the setting with <><> and <ENTER>.The minutes are highlighted.

Pagina 33

MultiLab 4010-3 pH valueba76143e01 04/2013 275 pH value5.1 Measuring5.1.1 Measuring the pH valueStability control(AutoRead)& HOLD functionThe stab

Pagina 34 - 34 ba76143e01 04/2013

pH value MultiLab 4010-328 ba76143e01 04/2013Criteria for a stablemeasured valueThe Stability control function checks whether the measured values are

Pagina 35

MultiLab 4010-3 pH valueba76143e01 04/2013 29The display of the temperature indicates the active temperature measuring mode:5.2 pH calibration5.2.1 Wh

Pagina 36 - _> key

MultiLab 4010-3 - ContentsMultiLab 4010-3 Contentsba76143e01 04/2013 31 Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 37 - _>, output all cali

pH value MultiLab 4010-330 ba76143e01 04/2013Continuing with two-point calibration4. Thoroughly rinse the sensor with deionized water.5. Immerse the s

Pagina 38 - FUNCTION

MultiLab 4010-3 pH valueba76143e01 04/2013 31Continuing withthree- to five-pointcalibration12. When measuring without temperature sensor:Enter the tem

Pagina 39

pH value MultiLab 4010-332 ba76143e01 04/20135.2.4 Carrying out a manual calibration (ConCal)Make sure that in the sensor menu, Buffer menu, the ConCa

Pagina 40 - 40 ba76143e01 04/2013

MultiLab 4010-3 pH valueba76143e01 04/2013 334. Thoroughly rinse the sensor with deionized water.5. Immerse the pH sensor in buffer solution 1.6. For

Pagina 41

pH value MultiLab 4010-334 ba76143e01 04/2013Continuing with two-point calibration9. Set the nominal buffer value for the measured temperature with &l

Pagina 42 - 42 ba76143e01 04/2013

MultiLab 4010-3 pH valueba76143e01 04/2013 35Continuing withthree- to five-pointcalibration5.2.5 Calibration pointsCalibration can be performed using

Pagina 43 - 6 ORP voltage

pH value MultiLab 4010-336 ba76143e01 04/20135.2.6 Calibration dataDisplaying the cali-bration dataThe calibration record of the last calibration is t

Pagina 44 - Δ : Better than 0.3 mV

MultiLab 4010-3 pH valueba76143e01 04/2013 37Calibration evalua-tionAfter calibrating, the meter automatically evaluates the calibration. The zero poi

Pagina 45 - 6.2 ORP calibration

pH value MultiLab 4010-338 ba76143e01 04/2013Calibration record5.2.7 Continuous measurement control (CMC function)The Continuous Measurement Control (

Pagina 46 - 7 Dissolved oxygen (D.O.)

MultiLab 4010-3 pH valueba76143e01 04/2013 39The following information is displayed:The limits of the calibrated range are determined by the buffers u

Pagina 47 - ATURATION LOCAL, page 68)

Contents MultiLab 4010-34 ba76143e01 04/20135 pH value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 275.1 Measu

Pagina 48 - 7.2 Calibration

pH value MultiLab 4010-340 ba76143e01 04/20135.2.8 QSC function (sensor quality control)General informationon the QSC functionThe QSC function (Qualit

Pagina 49

MultiLab 4010-3 pH valueba76143e01 04/2013 41Carry out the normal calibrations for your special measuring range with your usual standard solutions as

Pagina 50 - 50 ba76143e01 04/2013

pH value MultiLab 4010-342 ba76143e01 04/2013sor evaluation.Carrying out a QSCcontrol calibrationA QSC control calibration can, e.g. be useful if the

Pagina 51

MultiLab 4010-3 ORP voltageba76143e01 04/2013 436 ORP voltage6.1 Measuring6.1.1 Measuring the ORPStability control(AutoRead)& HOLD functionThe sta

Pagina 52 - 52 ba76143e01 04/2013

ORP voltage MultiLab 4010-344 ba76143e01 04/2013Criteria for a stablemeasured valueThe Stability control function checks whether the measured values a

Pagina 53 - _>, output all

MultiLab 4010-3 ORP voltageba76143e01 04/2013 45to measure and enter the temperature of the sample.The measuring instrument recognizes whether a suita

Pagina 54 - HAT TO DO IF...

Dissolved oxygen (D.O.) MultiLab 4010-346 ba76143e01 04/20137 Dissolved oxygen (D.O.)7.1 Measuring7.1.1 Measuring D.O.Selecting thedisplayedmeasured p

Pagina 55 - 8 Conductivity

MultiLab 4010-3 Dissolved oxygen (D.O.)ba76143e01 04/2013 47You can view the current air pressure in the sensor menu when an IDS D.O. sensor is connec

Pagina 56 - Δ ϰ: better than

Dissolved oxygen (D.O.) MultiLab 4010-348 ba76143e01 04/2013Criteria for a stablemeasured valueThe Stability control function checks whether the measu

Pagina 57 - 8.2 Temperature compensation

MultiLab 4010-3 Dissolved oxygen (D.O.)ba76143e01 04/2013 49 Calibration in water vapor-saturated air. Calibration via a comparison measurement (e.g

Pagina 58 - 8.3 Calibration

MultiLab 4010-3 Contentsba76143e01 04/2013 59.3 D.O. measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 669.3.1 Settings for

Pagina 59

Dissolved oxygen (D.O.) MultiLab 4010-350 ba76143e01 04/20137.2.5 2-point calibrationCalibration in water vapor-saturated air (Cal) and calibration vi

Pagina 60

MultiLab 4010-3 Dissolved oxygen (D.O.)ba76143e01 04/2013 513. Place the D.O. sensor in a solution that does not contain any dissolved oxygen.A soluti

Pagina 61 - 9Settings

Dissolved oxygen (D.O.) MultiLab 4010-352 ba76143e01 04/20137. Put the IDS D.O. sensor into the calibration medium. When calibrating via comparison m

Pagina 62 - 62 ba76143e01 04/2013

MultiLab 4010-3 Dissolved oxygen (D.O.)ba76143e01 04/2013 537.2.6 Calibration dataDisplaying thecalibration recordThe calibration record of the last c

Pagina 63

Dissolved oxygen (D.O.) MultiLab 4010-354 ba76143e01 04/2013ProOBODcalibrationevaluationCalibration recordDisplay Calibration record Relative slope+++

Pagina 64 - * 4.000

MultiLab 4010-3 Conductivityba76143e01 04/2013 558 Conductivity8.1 Measuring8.1.1 Measuring the conductivitySelecting thedisplayedmeasured parameterYo

Pagina 65 - 9.2 ORP measurement settings

Conductivity MultiLab 4010-356 ba76143e01 04/2013You can start the Stability control manually at any time, irrespective of the set-ting for automatic

Pagina 66 - 9.3 D.O. measurement settings

MultiLab 4010-3 Conductivityba76143e01 04/2013 578.1.2 Measuring the temperatureFor reproducible conductivity measurements, it is essential to measure

Pagina 67 - ATURATION LOCAL, page 68

Conductivity MultiLab 4010-358 ba76143e01 04/20138.3 Calibration8.3.1 Why calibrate?Aging slightly changes the cell constant, e. g. due to coatings.

Pagina 68 - 9.4 Cond measurement settings

MultiLab 4010-3 Conductivityba76143e01 04/2013 598.3.4 Calibration dataThe calibration data can be displayed and then output to the interface.Displayi

Pagina 69

Contents MultiLab 4010-36 ba76143e01 04/201314 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9414.1 Measuri

Pagina 70 - 70 ba76143e01 04/2013

Conductivity MultiLab 4010-360 ba76143e01 04/2013CalibrationevaluationAfter calibration, the meter automatically evaluates the current status of the c

Pagina 71 - _> key in the mea

MultiLab 4010-3 Settingsba76143e01 04/2013 619Settings9.1 pH measurement settings9.1.1 Settings for pH measurementsThe settings are made in the menu f

Pagina 72 - 72 ba76143e01 04/2013

Settings MultiLab 4010-362 ba76143e01 04/20139.1.2 Buffer sets for calibrationYou can use the buffer sets quoted in the table for an automatic calibra

Pagina 73 - 9.6 Reset

MultiLab 4010-3 Settingsba76143e01 04/2013 635 Merck 2 * 1.0006.0008.00013.00020 °C6 Merck 3 * 4.6606.8809.22020 °C7 Merck 4 * 2.0004.0007.00010.00020

Pagina 74 - 74 ba76143e01 04/2013

Settings MultiLab 4010-364 ba76143e01 04/2013* Brand names or trade names are trademarks of their respective owners protected by law.17 Beckman * 4.00

Pagina 75 - _> key in the

MultiLab 4010-3 Settingsba76143e01 04/2013 659.1.3 Calibration intervalThe calibration evaluation is displayed as a sensor symbol.After the QSC functi

Pagina 76

Settings MultiLab 4010-366 ba76143e01 04/20139.3 D.O. measurement settings9.3.1 Settings for D.O. measurement The settings are available in the menu f

Pagina 77

MultiLab 4010-3 Settingsba76143e01 04/2013 679.3.2 Enter Sensor Cap coefficientsSalinity from chan-nel▭▭▬▭▬▭▬▭▭Selects the channel (sensor), that pro-

Pagina 78

Settings MultiLab 4010-368 ba76143e01 04/20139.3.3 Saturation localIrrespective of the height or air pressure, the calibration value is set to 100%. S

Pagina 79 - 10.3 Measurement data storage

MultiLab 4010-3 Settingsba76143e01 04/2013 69Type Used measuring cellCal Measuring cells whose cell constant is determined by calibration in the KCl c

Pagina 80

MultiLab 4010-3 Overviewba76143e01 04/2013 71 Overview1.1 Meter MultiLab 4010-3The compact, digital precision meter MultiLab 4010-3 enables you to car

Pagina 81

Settings MultiLab 4010-370 ba76143e01 04/2013Setting menu of4320Menu item Possible settingExplanationCell constant 0.090 ... ... 0.100 ... 0.110 cm-1D

Pagina 82 - 11 Transmitting data

MultiLab 4010-3 Settingsba76143e01 04/2013 719.5 Sensor-independent settings9.5.1 SystemTo open the Storage & config menu, press the <ENTER_>

Pagina 83 - (USB-A interface (USB Host))

Settings MultiLab 4010-372 ba76143e01 04/20139.5.2 Data storageThis menu contains all functions to display, edit and erase stored measured values.9.5.

Pagina 84 - 84 ba76143e01 04/2013

MultiLab 4010-3 Settingsba76143e01 04/2013 739.6 ResetYou can reset (initialize) all sensor settings and sensor-independent settings separately from e

Pagina 85 - 11.8 BOD Analyst Pro

Settings MultiLab 4010-374 ba76143e01 04/2013Dissolved oxygen The following settings for D.O. measurements are reset to the default settings with the

Pagina 86 - 12.1 Maintenance

MultiLab 4010-3 Settingsba76143e01 04/2013 759.6.2 Resetting the system settingsThe following system settings can be reset to the default condition:Th

Pagina 87 - 12.4 Disposal

Data storage MultiLab 4010-376 ba76143e01 04/201310 Data storageYou can transmit measured values (datasets) to the data storage: Manual data storage

Pagina 88 - 88 ba76143e01 04/2013

MultiLab 4010-3 Data storageba76143e01 04/2013 77Settings You can configure the automatic data storage function with the following settings:1 Adjusted

Pagina 89 - 13 What to do if

Data storage MultiLab 4010-378 ba76143e01 04/2013Starting theautomatic storagefunctionTo start the automatic storage function, select continue with &l

Pagina 90 - 13.2 Dissolved oxygen

MultiLab 4010-3 Data storageba76143e01 04/2013 7910.3 Measurement data storage10.3.1 Editing the measurement data storageThe contents of the manual or

Pagina 91 - 13.3 Conductivity

Overview MultiLab 4010-38 ba76143e01 04/20131.2 Sensors1.2.1 IDS sensorsIDS sensors  support the automatic sensor recognition show only the settings

Pagina 92 - 13.4 General information

Data storage MultiLab 4010-380 ba76143e01 04/2013Display presentationof a datasetSample printoutQuitting the display To quit the display of stored mea

Pagina 93

MultiLab 4010-3 Data storageba76143e01 04/2013 81 Measured value of the connected sensor Measured temperature value of the connected sensor AutoRea

Pagina 94 - 14 Technical data

Transmitting data MultiLab 4010-382 ba76143e01 04/201311 Transmitting data 11.1 Outputting current measurement data11.2 Transmitting dataThe meter has

Pagina 95

MultiLab 4010-3 Transmitting databa76143e01 04/2013 8311.4 Connecting a USB memory device/USB printer (USB-A interface (USB Host))Connect the USB-A in

Pagina 96 - CERTIFICATE OF ANALYSIS

Transmitting data MultiLab 4010-384 ba76143e01 04/201311.5 Options for data transmission to the USB-B (PC) interface and the USB-A (USB printer) inter

Pagina 97

MultiLab 4010-3 Transmitting databa76143e01 04/2013 8511.6 Data transmission to the USB-A interface (USB memory device)Via the USB-A interface, you ca

Pagina 98 - 15 Firmware update

Maintenance, cleaning, disposal MultiLab 4010-386 ba76143e01 04/201312 Maintenance, cleaning, disposal12.1 Maintenance12.1.1 General maintenance activ

Pagina 99

MultiLab 4010-3 Maintenance, cleaning, disposalba76143e01 04/2013 8712.2 CleaningOccasionally wipe the outside of the measuring instrument with a damp

Pagina 100 - 16 Glossary

Maintenance, cleaning, disposal MultiLab 4010-388 ba76143e01 04/2013

Pagina 101 - General information

MultiLab 4010-3 What to do if...ba76143e01 04/2013 8913 What to do if...13.1 pHError messageOFL, UFLError message,ErrorMore information and instructio

Pagina 102 - 102 ba76143e01 04/2013

MultiLab 4010-3 Overviewba76143e01 04/2013 91.2.2 IDS adapter for analog pH sensorsWith the aid of an IDS adapter, you can also operate analog sensors

Pagina 103 - 17 Index

What to do if... MultiLab 4010-390 ba76143e01 04/2013No stable measuredvalue Obviously incorrectmeasured values13.2 Dissolved oxygenError message,OFLE

Pagina 104 - 104 ba76143e01 04/2013

MultiLab 4010-3 What to do if...ba76143e01 04/2013 9113.3 ConductivityError message,OFLError message,Error– Measured temperature value outside the ope

Pagina 105 - MultiLab 4010-3 Index

What to do if... MultiLab 4010-392 ba76143e01 04/201313.4 General informationSensor symbolflashesMeter does not reactto keystrokeYou want to knowwhich

Pagina 106 - 18 Appendix

MultiLab 4010-3 What to do if...ba76143e01 04/2013 93Data transmission toUSB printer does notworkError message,Memory error 1Time gets lostCause Remed

Pagina 107 - MultiLab 4010-3 Appendix

Technical data MultiLab 4010-394 ba76143e01 04/201314 Technical data 14.1 Measuring ranges, resolution, accuracy14.2 General dataMeasuring ranges,accu

Pagina 108 - 18.2 DO% Calibration values

MultiLab 4010-3 Technical databa76143e01 04/2013 95Parity NoneHandshake RTS/CTSCable length Max. 3 mUSB interface (host) Type USB 2.0USB-A (host), US

Pagina 109 - 19 Contact Information

Technical data MultiLab 4010-396 ba76143e01 04/2013Keypad (antibacterial)LSN 25/7181515/07/2004CERTIFICATE OF ANALYSISSample Ref No.:Date Received:The

Pagina 110

MultiLab 4010-3 Technical databa76143e01 04/2013 97LSN 26/3812324/10/2004CERTIFICATE OF ANALYSISSample Ref No.:Date Received:The microbiological resul

Pagina 111

Firmware update MultiLab 4010-398 ba76143e01 04/201315 Firmware update15.1 Firmware update for the meter MultiLab 4010-3You can find available firmwar

Pagina 112 - Xylem

MultiLab 4010-3 Firmware updateba76143e01 04/2013 9915.2 Firmware-Update for IDS SensorsWith the "Firmware Update" program and a PC you can