OPERATIONS MANUALba76143e01 04/2013MultiLab 4010-3pH/ORP/D.O./CondDIGITAL METER FOR IDS SENSORS4010-3
Safety MultiLab 4010-310 ba76143e01 04/20132Safety2.1 Safety information2.1.1 Safety information in the operating manualThis operating manual provides
Glossary MultiLab 4010-3100 ba76143e01 04/201316 GlossarypH/ORPConductivityAsymmetry see zero pointElectromotive force ofa combinationelectrodeThe ele
MultiLab 4010-3 Glossaryba76143e01 04/2013 101Dissolved oxygenGeneral informationSalinity The absolute salinity SA of seawater corresponds to the rela
Glossary MultiLab 4010-3102 ba76143e01 04/2013Measured value The measured value is the special value of a measured parameter to be determined. It is g
MultiLab 4010-3 Indexba76143e01 04/2013 10317 IndexAAir calibration beaker . . . . . . . . . . . . . . . . . 49AutoRead . . . . . . . . . . . . . .
Index MultiLab 4010-3104 ba76143e01 04/2013Transmitting data . . . . . . . . . . . . . . . . . . . .82Automatic . . . . . . . . . . . . . . . . . .
MultiLab 4010-3 Indexba76143e01 04/2013 105
Appendix MultiLab 4010-3106 ba76143e01 04/201318 Appendix18.1 Oxygen solubility tableSolubility of oxygen in mg/L in water exposed to water-saturated
MultiLab 4010-3 Appendixba76143e01 04/2013 10733.0 7.183 6.843 6.518 6.208 5.912 5.63134.0 7.065 6.732 6.415 6.111 5.822 5.54635.0 6.950 6.624 6.314 6
Appendix MultiLab 4010-3108 ba76143e01 04/201318.2 DO% Calibration valuesPressure Altitude Calibration valueInches Hg mm Hg kPa mbar Feet Meters % Sat
MultiLab 4010-3 Contact Information109ba76143e01 04/201319 Contact Information19.1 Ordering & Technical SupportWhen placing an order please have t
MultiLab 4010-3 Safetyba76143e01 04/2013 11Only the operation and running of the meter according to the instructions and technical specifications give
Contact Information MultiLab 4010-3110ba76143e01 04/2013
1) The tissue in plants that brings water upward from the roots;2) a leading global water technology company.We're 12,500 people unified in a com
Commissioning MultiLab 4010-312 ba76143e01 04/20133 Commissioning3.1 Scope of delivery MeterMultiLab 4010-3 USB cable (A plug on mini B plug) Powe
MultiLab 4010-3 Commissioningba76143e01 04/2013 133.3.1 Connecting the power packCAUTIONThe line voltage at the operating site must lie within the inp
Operation MultiLab 4010-314 ba76143e01 04/20134 Operation4.1 General operating principles4.1.1 KeypadIn this operating manual, keys are indicated by b
MultiLab 4010-3 Operationba76143e01 04/2013 154.1.2 DisplayExample (pH):4.1.3 Status information1 Status information (meter)2 Status information (sens
Operation MultiLab 4010-316 ba76143e01 04/20134.1.4 Instrument connectors1 IDS sensors: (pH, ORP, conductivity, D.O.)a) Channel 1b) Channel 2c) Channe
MultiLab 4010-3 Operationba76143e01 04/2013 174.1.5 Channel displayThe MultiLab 4010-3 manages the connected sensors and displays which sen-sor is plu
Operation MultiLab 4010-318 ba76143e01 04/20134.1.7 Display of several sensors in the measuring modeThe measured values of the connected sensors can b
MultiLab 4010-3 Operationba76143e01 04/2013 194.2 Switching on the meter4.3 Switching off the meter4.4 Login with user nameAfter activation of the use
2 ba76143e01 04/2013MultiLab 4010-3For the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs, OH
Operation MultiLab 4010-320 ba76143e01 04/2013Changing thepasswordIf the administrator has set up the access with password protection:2. Using <&g
MultiLab 4010-3 Operationba76143e01 04/2013 21Forgotten thepassword?Contact the administrator.4.5 Navigation4.5.1 Operating modes4.5.2 Measured value
Operation MultiLab 4010-322 ba76143e01 04/2013The selection is made with the <><> keys. The current selection is dis-played with a frame
MultiLab 4010-3 Operationba76143e01 04/2013 23 MessagesInformation is marked by the [i] symbol. It cannot be selected. Example:4.5.4 Example 1 on nav
Operation MultiLab 4010-324 ba76143e01 04/20133. Select the System submenu with <><>. The current selection is displayed with a frame.4.
MultiLab 4010-3 Operationba76143e01 04/2013 254.5.5 Example 2 on navigation: Setting the date and timeThe meter has a clock with a date function. The
Operation MultiLab 4010-326 ba76143e01 04/20134. Change and confirm the setting with <><> and <ENTER>.The minutes are highlighted.
MultiLab 4010-3 pH valueba76143e01 04/2013 275 pH value5.1 Measuring5.1.1 Measuring the pH valueStability control(AutoRead)& HOLD functionThe stab
pH value MultiLab 4010-328 ba76143e01 04/2013Criteria for a stablemeasured valueThe Stability control function checks whether the measured values are
MultiLab 4010-3 pH valueba76143e01 04/2013 29The display of the temperature indicates the active temperature measuring mode:5.2 pH calibration5.2.1 Wh
MultiLab 4010-3 - ContentsMultiLab 4010-3 Contentsba76143e01 04/2013 31 Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
pH value MultiLab 4010-330 ba76143e01 04/2013Continuing with two-point calibration4. Thoroughly rinse the sensor with deionized water.5. Immerse the s
MultiLab 4010-3 pH valueba76143e01 04/2013 31Continuing withthree- to five-pointcalibration12. When measuring without temperature sensor:Enter the tem
pH value MultiLab 4010-332 ba76143e01 04/20135.2.4 Carrying out a manual calibration (ConCal)Make sure that in the sensor menu, Buffer menu, the ConCa
MultiLab 4010-3 pH valueba76143e01 04/2013 334. Thoroughly rinse the sensor with deionized water.5. Immerse the pH sensor in buffer solution 1.6. For
pH value MultiLab 4010-334 ba76143e01 04/2013Continuing with two-point calibration9. Set the nominal buffer value for the measured temperature with &l
MultiLab 4010-3 pH valueba76143e01 04/2013 35Continuing withthree- to five-pointcalibration5.2.5 Calibration pointsCalibration can be performed using
pH value MultiLab 4010-336 ba76143e01 04/20135.2.6 Calibration dataDisplaying the cali-bration dataThe calibration record of the last calibration is t
MultiLab 4010-3 pH valueba76143e01 04/2013 37Calibration evalua-tionAfter calibrating, the meter automatically evaluates the calibration. The zero poi
pH value MultiLab 4010-338 ba76143e01 04/2013Calibration record5.2.7 Continuous measurement control (CMC function)The Continuous Measurement Control (
MultiLab 4010-3 pH valueba76143e01 04/2013 39The following information is displayed:The limits of the calibrated range are determined by the buffers u
Contents MultiLab 4010-34 ba76143e01 04/20135 pH value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 275.1 Measu
pH value MultiLab 4010-340 ba76143e01 04/20135.2.8 QSC function (sensor quality control)General informationon the QSC functionThe QSC function (Qualit
MultiLab 4010-3 pH valueba76143e01 04/2013 41Carry out the normal calibrations for your special measuring range with your usual standard solutions as
pH value MultiLab 4010-342 ba76143e01 04/2013sor evaluation.Carrying out a QSCcontrol calibrationA QSC control calibration can, e.g. be useful if the
MultiLab 4010-3 ORP voltageba76143e01 04/2013 436 ORP voltage6.1 Measuring6.1.1 Measuring the ORPStability control(AutoRead)& HOLD functionThe sta
ORP voltage MultiLab 4010-344 ba76143e01 04/2013Criteria for a stablemeasured valueThe Stability control function checks whether the measured values a
MultiLab 4010-3 ORP voltageba76143e01 04/2013 45to measure and enter the temperature of the sample.The measuring instrument recognizes whether a suita
Dissolved oxygen (D.O.) MultiLab 4010-346 ba76143e01 04/20137 Dissolved oxygen (D.O.)7.1 Measuring7.1.1 Measuring D.O.Selecting thedisplayedmeasured p
MultiLab 4010-3 Dissolved oxygen (D.O.)ba76143e01 04/2013 47You can view the current air pressure in the sensor menu when an IDS D.O. sensor is connec
Dissolved oxygen (D.O.) MultiLab 4010-348 ba76143e01 04/2013Criteria for a stablemeasured valueThe Stability control function checks whether the measu
MultiLab 4010-3 Dissolved oxygen (D.O.)ba76143e01 04/2013 49 Calibration in water vapor-saturated air. Calibration via a comparison measurement (e.g
MultiLab 4010-3 Contentsba76143e01 04/2013 59.3 D.O. measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 669.3.1 Settings for
Dissolved oxygen (D.O.) MultiLab 4010-350 ba76143e01 04/20137.2.5 2-point calibrationCalibration in water vapor-saturated air (Cal) and calibration vi
MultiLab 4010-3 Dissolved oxygen (D.O.)ba76143e01 04/2013 513. Place the D.O. sensor in a solution that does not contain any dissolved oxygen.A soluti
Dissolved oxygen (D.O.) MultiLab 4010-352 ba76143e01 04/20137. Put the IDS D.O. sensor into the calibration medium. When calibrating via comparison m
MultiLab 4010-3 Dissolved oxygen (D.O.)ba76143e01 04/2013 537.2.6 Calibration dataDisplaying thecalibration recordThe calibration record of the last c
Dissolved oxygen (D.O.) MultiLab 4010-354 ba76143e01 04/2013ProOBODcalibrationevaluationCalibration recordDisplay Calibration record Relative slope+++
MultiLab 4010-3 Conductivityba76143e01 04/2013 558 Conductivity8.1 Measuring8.1.1 Measuring the conductivitySelecting thedisplayedmeasured parameterYo
Conductivity MultiLab 4010-356 ba76143e01 04/2013You can start the Stability control manually at any time, irrespective of the set-ting for automatic
MultiLab 4010-3 Conductivityba76143e01 04/2013 578.1.2 Measuring the temperatureFor reproducible conductivity measurements, it is essential to measure
Conductivity MultiLab 4010-358 ba76143e01 04/20138.3 Calibration8.3.1 Why calibrate?Aging slightly changes the cell constant, e. g. due to coatings.
MultiLab 4010-3 Conductivityba76143e01 04/2013 598.3.4 Calibration dataThe calibration data can be displayed and then output to the interface.Displayi
Contents MultiLab 4010-36 ba76143e01 04/201314 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9414.1 Measuri
Conductivity MultiLab 4010-360 ba76143e01 04/2013CalibrationevaluationAfter calibration, the meter automatically evaluates the current status of the c
MultiLab 4010-3 Settingsba76143e01 04/2013 619Settings9.1 pH measurement settings9.1.1 Settings for pH measurementsThe settings are made in the menu f
Settings MultiLab 4010-362 ba76143e01 04/20139.1.2 Buffer sets for calibrationYou can use the buffer sets quoted in the table for an automatic calibra
MultiLab 4010-3 Settingsba76143e01 04/2013 635 Merck 2 * 1.0006.0008.00013.00020 °C6 Merck 3 * 4.6606.8809.22020 °C7 Merck 4 * 2.0004.0007.00010.00020
Settings MultiLab 4010-364 ba76143e01 04/2013* Brand names or trade names are trademarks of their respective owners protected by law.17 Beckman * 4.00
MultiLab 4010-3 Settingsba76143e01 04/2013 659.1.3 Calibration intervalThe calibration evaluation is displayed as a sensor symbol.After the QSC functi
Settings MultiLab 4010-366 ba76143e01 04/20139.3 D.O. measurement settings9.3.1 Settings for D.O. measurement The settings are available in the menu f
MultiLab 4010-3 Settingsba76143e01 04/2013 679.3.2 Enter Sensor Cap coefficientsSalinity from chan-nel▭▭▬▭▬▭▬▭▭Selects the channel (sensor), that pro-
Settings MultiLab 4010-368 ba76143e01 04/20139.3.3 Saturation localIrrespective of the height or air pressure, the calibration value is set to 100%. S
MultiLab 4010-3 Settingsba76143e01 04/2013 69Type Used measuring cellCal Measuring cells whose cell constant is determined by calibration in the KCl c
MultiLab 4010-3 Overviewba76143e01 04/2013 71 Overview1.1 Meter MultiLab 4010-3The compact, digital precision meter MultiLab 4010-3 enables you to car
Settings MultiLab 4010-370 ba76143e01 04/2013Setting menu of4320Menu item Possible settingExplanationCell constant 0.090 ... ... 0.100 ... 0.110 cm-1D
MultiLab 4010-3 Settingsba76143e01 04/2013 719.5 Sensor-independent settings9.5.1 SystemTo open the Storage & config menu, press the <ENTER_>
Settings MultiLab 4010-372 ba76143e01 04/20139.5.2 Data storageThis menu contains all functions to display, edit and erase stored measured values.9.5.
MultiLab 4010-3 Settingsba76143e01 04/2013 739.6 ResetYou can reset (initialize) all sensor settings and sensor-independent settings separately from e
Settings MultiLab 4010-374 ba76143e01 04/2013Dissolved oxygen The following settings for D.O. measurements are reset to the default settings with the
MultiLab 4010-3 Settingsba76143e01 04/2013 759.6.2 Resetting the system settingsThe following system settings can be reset to the default condition:Th
Data storage MultiLab 4010-376 ba76143e01 04/201310 Data storageYou can transmit measured values (datasets) to the data storage: Manual data storage
MultiLab 4010-3 Data storageba76143e01 04/2013 77Settings You can configure the automatic data storage function with the following settings:1 Adjusted
Data storage MultiLab 4010-378 ba76143e01 04/2013Starting theautomatic storagefunctionTo start the automatic storage function, select continue with &l
MultiLab 4010-3 Data storageba76143e01 04/2013 7910.3 Measurement data storage10.3.1 Editing the measurement data storageThe contents of the manual or
Overview MultiLab 4010-38 ba76143e01 04/20131.2 Sensors1.2.1 IDS sensorsIDS sensors support the automatic sensor recognition show only the settings
Data storage MultiLab 4010-380 ba76143e01 04/2013Display presentationof a datasetSample printoutQuitting the display To quit the display of stored mea
MultiLab 4010-3 Data storageba76143e01 04/2013 81 Measured value of the connected sensor Measured temperature value of the connected sensor AutoRea
Transmitting data MultiLab 4010-382 ba76143e01 04/201311 Transmitting data 11.1 Outputting current measurement data11.2 Transmitting dataThe meter has
MultiLab 4010-3 Transmitting databa76143e01 04/2013 8311.4 Connecting a USB memory device/USB printer (USB-A interface (USB Host))Connect the USB-A in
Transmitting data MultiLab 4010-384 ba76143e01 04/201311.5 Options for data transmission to the USB-B (PC) interface and the USB-A (USB printer) inter
MultiLab 4010-3 Transmitting databa76143e01 04/2013 8511.6 Data transmission to the USB-A interface (USB memory device)Via the USB-A interface, you ca
Maintenance, cleaning, disposal MultiLab 4010-386 ba76143e01 04/201312 Maintenance, cleaning, disposal12.1 Maintenance12.1.1 General maintenance activ
MultiLab 4010-3 Maintenance, cleaning, disposalba76143e01 04/2013 8712.2 CleaningOccasionally wipe the outside of the measuring instrument with a damp
Maintenance, cleaning, disposal MultiLab 4010-388 ba76143e01 04/2013
MultiLab 4010-3 What to do if...ba76143e01 04/2013 8913 What to do if...13.1 pHError messageOFL, UFLError message,ErrorMore information and instructio
MultiLab 4010-3 Overviewba76143e01 04/2013 91.2.2 IDS adapter for analog pH sensorsWith the aid of an IDS adapter, you can also operate analog sensors
What to do if... MultiLab 4010-390 ba76143e01 04/2013No stable measuredvalue Obviously incorrectmeasured values13.2 Dissolved oxygenError message,OFLE
MultiLab 4010-3 What to do if...ba76143e01 04/2013 9113.3 ConductivityError message,OFLError message,Error– Measured temperature value outside the ope
What to do if... MultiLab 4010-392 ba76143e01 04/201313.4 General informationSensor symbolflashesMeter does not reactto keystrokeYou want to knowwhich
MultiLab 4010-3 What to do if...ba76143e01 04/2013 93Data transmission toUSB printer does notworkError message,Memory error 1Time gets lostCause Remed
Technical data MultiLab 4010-394 ba76143e01 04/201314 Technical data 14.1 Measuring ranges, resolution, accuracy14.2 General dataMeasuring ranges,accu
MultiLab 4010-3 Technical databa76143e01 04/2013 95Parity NoneHandshake RTS/CTSCable length Max. 3 mUSB interface (host) Type USB 2.0USB-A (host), US
Technical data MultiLab 4010-396 ba76143e01 04/2013Keypad (antibacterial)LSN 25/7181515/07/2004CERTIFICATE OF ANALYSISSample Ref No.:Date Received:The
MultiLab 4010-3 Technical databa76143e01 04/2013 97LSN 26/3812324/10/2004CERTIFICATE OF ANALYSISSample Ref No.:Date Received:The microbiological resul
Firmware update MultiLab 4010-398 ba76143e01 04/201315 Firmware update15.1 Firmware update for the meter MultiLab 4010-3You can find available firmwar
MultiLab 4010-3 Firmware updateba76143e01 04/2013 9915.2 Firmware-Update for IDS SensorsWith the "Firmware Update" program and a PC you can