Xylem 6-Series Multiparameter Manual de usuario Pagina 1

Busca en linea o descarga Manual de usuario para Herramientas Xylem 6-Series Multiparameter. Xylem 6-Series Multiparameter User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 377
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente

Indice de contenidos

Pagina 1 - User Manual

6-Series Multiparameter Water Quality Sondes User Manual 6-Series: 6600 V2 6600EDS V2 6920 V2 682

Pagina 2 - SAFETY NOTES

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-4 Figure 4 Sonde to 650 Display/Logger You will need

Pagina 3 - TABLE OF CONTENTS

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-94 You may choose the entire file. Use the Space Bar to a

Pagina 4

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-95 2.9.4 STATUS Select 4-Status from the Sonde Main menu

Pagina 5 - 1.2 HOW TO USE THIS MANUAL

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-96 2.9.5 SYSTEM Select 5-System from the Sonde Main menu

Pagina 6 - Introduction Section 1

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-97 The default is 9600, but you may change it to match you

Pagina 7 - 2.1 GETTING STARTED

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-98 The asterisks (*) that follow the numbers or letters i

Pagina 8 - 2.2 CONNECTING YOUR SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-99 Oxygen in % air saturation, Dissolved Oxygen in mg/L, p

Pagina 9 - You will need

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-100 PAR1 Output from special photosynthetically active ra

Pagina 10 - Sonde to 650 Display/Logger

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-101 As noted above, the ISE3 PAR1 selection is used in a s

Pagina 11 - DRY MEMBRANE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-102 Select 1-Cal constants to display the calibration con

Pagina 12

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-103 Turb+ A2 1000 0.6 to 1.5 Range is ratio of (M2

Pagina 13

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-5 2.3 PREPARING THE SONDE FOR USE To prepare the sonde

Pagina 14 - PROBE PORT PLUGS

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-104 Users should select the option for their

Pagina 15 - ISE PORT PLUG

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-105 6-( ) Multi SDI12. Modifies the SDI12 protocol as fo

Pagina 16

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-106 NOTE: The number of items on this menu depends great

Pagina 17 - Figure 20C

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-107 calculation of depth or level. This item will only app

Pagina 18

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-108 activated at the interval assigned in the Unattended s

Pagina 19

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-109 Sondes with no optical (turbidity, chlorophyll, or rho

Pagina 20 - LUBRICATE O-RINGS

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-110 Choosing 4-Threshold will display the following:

Pagina 21 - OPTIC

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-111 reading in air is near zero, and quite likely, is a ve

Pagina 22 - PROBE GUARD

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-112 When caring for your sonde, remember that the sonde is

Pagina 23 - 2.3.3 STEP 3 - POWER

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-113 SONDE PROBE PORTS Whenever you install, remove or r

Pagina 24 -

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-6 DO MEMBRANE INSTALLATION WITH THE PROBE NOT INSTALLED

Pagina 25

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-114 To resurface the probe using the fine sanding disk, f

Pagina 26

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-115 YSI recommends the optical DO membrane assembly be re

Pagina 27 - WITH CAP

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-116 Membrane Installation Instructions Use the schematic

Pagina 28

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-117  Store the probe with new membrane in either water o

Pagina 29

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-118 2. Rinse the probe in clean tap water, wipe with a co

Pagina 30

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-119 regeneration cycle has been successful. The felt filt

Pagina 31 - 2.5 SONDE SOFTWARE SETUP

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-120 2.10.3 SONDE STORAGE Proper storage of the your s

Pagina 32

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-121 plug. Leave the conductivity/temperature and the diss

Pagina 33

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-122 RAPID PULSE DISSOLVED OXYGEN Rapid Pulse dissolved

Pagina 34 - SONDE MENU FLOW CHART

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-123 ROX OPTICAL DISSOLVED OXYGEN When the 6150 sensor i

Pagina 35

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-7 DO MEMBRANE INSTALLATION WITH THE PROBE INSTALLED IN THE

Pagina 36

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-124 AMMONIUM, NITRATE AND CHLORIDE The active element in

Pagina 37

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-1 SECTION 3 650 MDS DISPLAY/LOGGER 3.1 INTRODUCTION The YSI 650 M

Pagina 38

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-2 3.2.1 UNPACKING Remove the instrument from the shipping box, being

Pagina 39

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-3 used to log data in Discrete or Unattended sampling applications.

Pagina 40 - 2.6.1 INTRODUCTION

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-4 Figure 2. 650 interfaced to a 6-series sonde with user-supplied GPS

Pagina 41 - USING THE CALIBRATION CUP

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-5 Figure 4. Charging of 650 rechargeable battery pack. The setup o

Pagina 42

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-6 3.2.4 650 FEATURES The key physical features of the 650 display a

Pagina 43

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-7 Note that the back of the case contains the battery lid that is a

Pagina 44

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-8 The circuit board in this instrument contains a manganese dioxide li

Pagina 45

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-9  Make certain that the gasket is properly installed on the rechar

Pagina 46

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-8 your fingers. Once the O-ring is in position, squeeze i

Pagina 47 - Sampling Applications

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-10 To use the device with an American or Japanese vehicle, slide th

Pagina 48

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-11 Note that the screen is divided into two sections – the “Main Displ

Pagina 49

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-12 3.2.8 USING THE YSI 650 KEYPAD The

Pagina 50

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-13 1 -- 1 2 – ABC2abc2 3 – DEF3def3 4 – GHI4ghi4 5 – JKL5jkl5 6 – MNO6

Pagina 51 - TURBIDITY 2-POINT

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-14 Use the arrow keys to highlight the Date and Time selection and pre

Pagina 52

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-15 As shown above, enter the designation “

Pagina 53

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-16 “click”. After connecting the sonde to the 650, press the “on” but

Pagina 54 - RHODAMINE WT 2-POINT

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-17 Section 3.11 of this manual. If symptoms persist after these acti

Pagina 55 - 2.7 TAKING READINGS

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-18 3.2.11 UPGRADING ECOWATCH FOR WINDOWS SOFTWARE

Pagina 56

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-19 YSI suggests that the user become familiar with th

Pagina 57

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-9 REMOVING THE PORT PLUGS Using the long extended end of

Pagina 58

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-20 3.3.7 ID Activate this item and Enter choice of instrument identi

Pagina 59

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-21 3.4 SONDE MENU INTERFACE 3.4.1 INTRODUCTION The Sonde menu se

Pagina 60 - ANALYZE DATA

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-22 In fact, the only significant difference between the two interfac

Pagina 61

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-23 3.4.2 SONDE MENU EXAMPLES To become familiar with the 650 Sonde

Pagina 62

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-24 Use the arrow keys to highlight the Dissolved Oxyge

Pagina 63 - OPENING A DATA FILE

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-25 “Calibrate” to “Continue”. Make sure that “Continue” is highlighte

Pagina 64 - VIEWING DATA

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-26 3.4.3.4 Calibrating sonde sensors Use the 650 Sonde menu selection

Pagina 65

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-27 The rules are particularly important with regard to the method of d

Pagina 66

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-28 The user should also

Pagina 67

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-29  Decide if you want to log data continuously, and, if so, determ

Pagina 68

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-10 600XL & 600XLM SONDE BULKHEAD 3 Port Sonde: 1 Rapi

Pagina 69

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-30 The user places the sonde in the water and watches the readings on

Pagina 70

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-31 The user attaches the 6920 sonde to a 650. After turning on the 6

Pagina 71

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-32 At the end of the 90 minute profile determination, the

Pagina 72

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-33 3.5.3 LOGGING DATA TO 650 MEMORY 3.5.3.1 INTRODUCTION Even if

Pagina 73

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-34 Press Enter at the highlighted Interval selection and use the arrow

Pagina 74 - 2.9 SONDE MENU

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-35 Using the arrow keys, the Filename window is highlighted, the user

Pagina 75 - 2.9.1 RUN

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-36 3.5.3.3 LOGGING TO 650 MEMORY WITH A SITE LIST In the example abov

Pagina 76

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-37 Note that Site lists containing Single Site Des

Pagina 77 - UNATTENDED SAMPLING

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-38 Note also that if you have purchased a 650 without the b

Pagina 78

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-39 Then press Enter. When the logging prompt is activated,

Pagina 79

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-11 6600EDS V2-2 SONDE BULKHEAD 5 Port Sonde: 1 Rapid Pulse

Pagina 80

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-40 These files will remain “open” in 650 memory so that yo

Pagina 81 - 2.9.2 CALIBRATE

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-41 Then highlight the Edit site list selection and press

Pagina 82

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-42 Then move to the next location (EAST), activate the 650 logging fun

Pagina 83 - ROX OPTICAL DISSOLVED OXYGEN

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-43 Note that the data from two sites is identified by the Site Numbe

Pagina 84

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-44 The two inserted entries can be moved either up or down

Pagina 85

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-45 3.5.4 650 LOGGING – “CANS” AND “CAN’TS” The use of the 650 to fa

Pagina 86

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-46 3.5.4.2 SUMMARY OF 650 LOGGING LIMITATIONS WITH A 650/YSI SONDE SY

Pagina 87 - PRESSURE – ABS AND GAGE

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-47 To proceed with the details of 650 File management, turn the instru

Pagina 88 - AMMONIUM

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-48 For both files, the file name, site name, file size and time of th

Pagina 89 - NITRATE

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-49 3.6.3 UPLOAD TO PC This frequently used command is used to trans

Pagina 90 - OPTIC TURBIDITY

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-12 6820V2-1 & 6920V2-1 SONDE BULKHEADS 7 Por

Pagina 91 - OPTIC CHLOROPHYLL

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-50 To transfer data from the 650 to your PC, highlight the Upload to P

Pagina 92 - OPTIC BGA-PC

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-51 Use the arrow keys to scroll vertically to view more data points

Pagina 93 - OPTIC BGA-PE

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-52 3.7 UPLOADING DATA FROM SONDES 3.7.1 INTRODUCTION A primary f

Pagina 94

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-53 3.8 USING GPS WITH THE 650 3.8.1 SETTING UP THE GPS-650 INTERF

Pagina 95

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-54  The GPS readings must be logged to 650 memory from the 650 Run d

Pagina 96

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-55 3.9 USING THE 650 BAROMETER 3.9.1 PRINCIPLES OF OPERATION The

Pagina 97

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-56 3.9.4 LOGGING BAROMETER READINGS To log barometric

Pagina 98 - 2.9.3 FILE

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-57 3.10 UPGRADING 650 SOFTWARE The 650 software can easily be upgra

Pagina 99

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-58 3.11 TROUBLESHOOTING The following sections describes problems w

Pagina 100

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-59 3.12 FERRITE BEAD INSTALLATION WARNING: If you are using your 65

Pagina 101 - 2.9.4 STATUS

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-13 600LS BULKHEAD Fi

Pagina 102 - 2.9.5 SYSTEM

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-60 3.13 650 SAFETY CONSIDERATIONS IMPORTANT SAFETY INSTRUCTIONS! S

Pagina 103 - Sondes Section 2

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-61 Danger Notifications – Misuse creates a STRONG possibility of death

Pagina 104

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-62 616 Cigarette Lighter Charger Safety Information 1. This section

Pagina 105

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-63 CAUTION: If water has contacted the rechargeable battery pack, do n

Pagina 106 - 2.9.7 SENSOR

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-64 Barometer: Range: 500 to 800 mm Hg Resolution: 0.1 mm Hg Accuracy:

Pagina 107

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-65

Pagina 108

650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-66

Pagina 109

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-1 SECTION 4 ECOWATCH FOR WINDOWS 4.1 INTRODUCTIO

Pagina 110

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-2 6 - Page The Data Display Window can be divided in

Pagina 111

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-3 EcoWatch starts without an open data file (or .DAT fi

Pagina 112

SAFETY NOTES TECHNICAL SUPPORT AND WARRANTY INFORMATION Contact information for technical support and warranty information on YSI’s Environmental

Pagina 113

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-14 600 OMS V2-1 BULKHEAD

Pagina 114

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-4 Click To: Open an existing data file (.DAT). Eco

Pagina 115

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-5 4.1.5 TYPICAL APPLICATION Suppose you want to mea

Pagina 116

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-6 5. Print Your Graph To print the graph, open the F

Pagina 117

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-7 4.2 DATA ACQUISTION AND ANALYSIS 4.2.1 CONNECT A Y

Pagina 118 - O-RING CARE AND MAINTENANCE

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-8 A sonde or 650 display/logger must be connected to o

Pagina 119 - 6562 RAPID PULSE DO PROBES

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-9 To transfer data from the 650 to your PC, highlight

Pagina 120

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-10 Unzoom Use this button to view the study from

Pagina 121

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-11 4.2.5 GRAPH DATA IN REAL TIME Choosing the New co

Pagina 122

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-12 4.3 ECOWATCH MENU 4.3.1 FILE The File menu off

Pagina 123

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-13 CLOSE Use this command to close all windows contai

Pagina 124 - LEVEL SENSOR

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-15 INSTALLING THE TURBIDITY, CHLOROPHYLL, RHODAMINE WT

Pagina 125 - NITRATE PROBE

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-14 COPY TO CLIPBOARD If a graph or table is active,

Pagina 126 - 2.10.3 SONDE STORAGE

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-15 SAVE DATA DISPLAY When opening a data file, you wil

Pagina 127 - 2.10.4 PROBE STORAGE

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-16 EXPORT Use this command to send the current dat

Pagina 128

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-17 individual data entries, quotation marks surround an

Pagina 129

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-18 PRINT Use this command to print a document. If a gr

Pagina 130

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-19 PRINT SETUP Use this command to select a printer

Pagina 131 - 3.2 GETTING STARTED

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-20 Orientation Choose Portrait or Landscape. Paper Siz

Pagina 132 - 3.2.1 UNPACKING

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-21 4.3.2 EDIT The edit menu appears whenever a data

Pagina 133 - 3.2.3 650 CONFIGURATIONS

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-22 INSERT PARAMETER Adds an additional parameter to th

Pagina 134 - 650 MDS Section 3

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-23 NUM The Num Lock key is latched down. SCRL T

Pagina 135

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-16 INSTALLING THE ISE PROBES Fi

Pagina 136 - 3.2.4 650 FEATURES

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-24 Shortcuts Toolbar: STUDY INFO This command wi

Pagina 137

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-25 ZOOM IN Enlarges graph or data table (whichever i

Pagina 138 - (4) Screws

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-26 Port Setup This is where you can set port param

Pagina 139

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-27 If a modem is selected for a particular COM port,

Pagina 140

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-28  Dialing If there is no entry for Phone Number, th

Pagina 141

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-29  Default Sonde Port This refers to the COM port yo

Pagina 142

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-30 A common example is using a sonde in Run Mode. The

Pagina 143

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-31 4.3.5 REAL-TIME The Real-time menu offers the fo

Pagina 144

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-32 4.3.6 GRAPH ZOOM WINDOW Enlarges a section of

Pagina 145

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-33 MANUAL SCALE Takes you to the Graph Y-Axis dialog

Pagina 146

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-17 2.3.3 STEP 3 - POWER Some type of external power su

Pagina 147

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-34 Add/Remove There are two lists, Selected Parameter

Pagina 148 - 3.3 SETTING UP THE 650

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-35 Attributes This command takes you to the Change

Pagina 149

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-36 Calibrate TSS TSS is an abbreviation for Total

Pagina 150

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-37 Once you have at least two correlation points, you h

Pagina 151 - 3.4 SONDE MENU INTERFACE

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-38  Color Type or select one of the 16 predefined col

Pagina 152

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-39 Note that there are two color buttons, one for the t

Pagina 153 - 3.4.2 SONDE MENU EXAMPLES

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-40 1 - Open a data file and observe the presentation

Pagina 154

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-41 4.3.9 WINDOW The Window menu offers the follow

Pagina 155 - SELECTION

EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-42 4.3.10 HELP The Help menu offers the following co

Pagina 156

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-1 SECTION 5 PRINCIPLES OF OPERATION 5.1 CONDUCTIVI

Pagina 157

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-18 INSTALLING BATTERIES The 600XLM, 6600V2-2, 6600EDS V

Pagina 158

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-2 4 - Perform sensor calibration at a temperature a

Pagina 159 - 3.5.2.1 INTRODUCTION

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-3 generally speaking, TDS cannot be calculated accur

Pagina 160 - MEMORY

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-4 MEASUREMENT AND CALIBRATION PRECAUTIONS 1 - Ins

Pagina 161

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-5 CALIBRATION AND EFFECT OF TEMPERATURE The softw

Pagina 162

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-6 4 - During pH calibration, allow the sensors time

Pagina 163 - 3.5.3.1 INTRODUCTION

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-7 MEASUREMENT AND CALIBRATION PRECAUTIONS (1) Be ce

Pagina 164

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-8 Teflon membrane. The current associated with this

Pagina 165

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-9 protocol according to the instructions. The calib

Pagina 166

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-10 (4) For short term storage (2 weeks or less), k

Pagina 167

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-11 EFFECT OF TEMPERATURE

Pagina 168

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-19 INSTALLING BATTERIES INTO THE YSI 6600V2-2, 6600EDS V2-

Pagina 169

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-12 MEASUREMENT AND CALIBRATION PRECAUTIONS (1) If

Pagina 170

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-13 change in voltage. Our experience with ISEs indi

Pagina 171

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-14 use the less time-consuming 2 point and 1-point r

Pagina 172

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-15 = 0.6). Even if the ionic content were due to so

Pagina 173

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-16 (2) Ion selective electrodes may not stabilize as

Pagina 174

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-17 instrument can be frequently calibrated. However

Pagina 175

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-18 sonde software to provide readings in nephelometr

Pagina 176 - 3.6 MANAGING 650 FILES

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-19 turbidity standards supplied by YSI, the value of

Pagina 177 - 3.6.2 DIRECTORY

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-20 example, the label of the YSI 6073 turbidity stan

Pagina 178

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-21 For most unattended sampling applications, select

Pagina 179 - 3.6.3 UPLOAD TO PC

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-20 2. Press two of the outermost retaining tabs down into

Pagina 180 - 3.6.4 VIEW FILE

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-22 chlorophyll system are designed to complement the

Pagina 181 - 3.6.6 DELETE ALL FILES

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-23 6025 chlorophyll probe is equipped with a mechani

Pagina 182 - 3.7.2 UPLOAD PROCEDURE

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-24 PREPARATION OF DYE SOLUTIONS FOR CHECKING SENSOR

Pagina 183 - 3.8 USING GPS WITH THE 650

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-25 Table. Approximate algal chlorophyll equivalent o

Pagina 184

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-26 wiper that makes it ideal for unattended applicat

Pagina 185

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-27 standard that provides the best measure of accura

Pagina 186

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-28 content of an environmental sample may appear to

Pagina 187

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-29 Field optical measurements are particularly su

Pagina 188 - 3.11 TROUBLESHOOTING

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-30 RHODAMINE WT STANDARD SOLUTION – PREPARATION AND

Pagina 189

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-31 experiments indicate that a suspension of phytopl

Pagina 190 - Restrictions on Usage

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-21 cannot be reset and the battery lid must be replaced be

Pagina 191

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-32 described above. The YSI 6131 sensor is designed

Pagina 192

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-33 CALIBRATION METHODS For calibration of the YSI B

Pagina 193

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-34 2. Accurately weigh 0.500 g of the 20 % Rhodamin

Pagina 194

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-35 EFFECT OF CHLOROPHYLL ON BGA-PC READINGS While t

Pagina 195

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-36 capability of mechanical cleaning is likely to be

Pagina 196

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-37 5.17 PHYCOERYTHRIN-CONTAINING BLUE-GREEN ALGAE

Pagina 197 - 4.1.2 GLOSSARY

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-38 Measurement of PE-containing BGA In Vivo One ke

Pagina 198 - 4.1.3 TUTORIAL

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-39 CALIBRATION METHODS For calibration of the YSI B

Pagina 199 - 4.1.4 USING THE TOOLBAR

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-40 2. Accurately weigh 0.500 g of the 20 % Rhodamin

Pagina 200

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-41 EFFECT OF CHLOROPHYLL ON BGA-PE READINGS While

Pagina 201 - 4.1.5 TYPICAL APPLICATION

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-22 Check the O-rings and sealing surfaces for any contamin

Pagina 202

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-42 mechanical wiper that makes it ideal for unattend

Pagina 203 - 650 MDS to Lab Computer

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-43 n 5.18 FLOW Flow is a calculated value. Whene

Pagina 204

Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-44 vegetation in the channel also affects the value

Pagina 205 - 4.2.4 USING THE GRAPH

Troubleshooting Section 6 YSI Incorporated Environmental Monitoring Systems Operations Manual 6-1 SECTION 6 TROUBLESHOOTING This section co

Pagina 206

Troubleshooting Section 6 YSI Incorporated Environmental Monitoring Systems Operations Manual 6-2 6.2 SONDE COMMUNICATION PROBLEMS SYMPTOM POSSI

Pagina 207

Troubleshooting Section 6 YSI Incorporated Environmental Monitoring Systems Operations Manual 6-3 6.3 SENSOR PERFORMANCE PROBLEMS SYMPTOMS POSSI

Pagina 208 - 4.3 ECOWATCH MENU

Troubleshooting Section 6 YSI Incorporated Environmental Monitoring Systems Operations Manual 6-4 Water in probe connector Dry connector; reinst

Pagina 209

Communications Section 7 YSI Incorporated Environmental Monitoring Systems Operations Manual 7-1 SECTION 7 COMMUNICATION This section describes

Pagina 210 - COPY TO CLIPBOARD

Communications Section 7 YSI Incorporated Environmental Monitoring Systems Operations Manual 7-2 7.3 RS-232 INTERFACE The sonde has an auto-baud

Pagina 211 - LOAD DATA DISPLAY

Communications Section 7 YSI Incorporated Environmental Monitoring Systems Operations Manual 7-3 ttt - Maximum time in seconds the Sonde will ta

Pagina 212

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-23 CAUTION: The 6067B cable is for laboratory use only --

Pagina 213

Communications Section 7 YSI Incorporated Environmental Monitoring Systems Operations Manual 7-4 Manufacturer: YSIIWQSG Model: EMS60

Pagina 214

Upgrading Sonde Firmware Section 8 YSI Incorporated Environmental Monitoring Systems Operations Manual 8-1 SECTIO

Pagina 215 - PRINT SETUP

Upgrading Sonde Firmware Section 8 YSI Incorporated Environmental Monitoring Systems Operations Manual 8-2

Pagina 216

Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-1

Pagina 217 - 4.3.2 EDIT

Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-2

Pagina 218 - 4.3.3 VIEW

Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-3

Pagina 219

Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-4

Pagina 220 - STUDY INFO

Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-5

Pagina 221 - 4.3.4 COMM(UNICATION)

Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-1 APPENDIX A HEALTH AND SAFETY NOTE: For a

Pagina 222

Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-2 pH 4 INGREDIENTS: θ Potassium Hyd

Pagina 223

TABLE OF CONTENTS SECTION 1 INTRODUCTION 1.1 ABOUT YSI 1-1 1.2 HOW TO USE THIS MANUAL 1-1 1.3 UNPAC

Pagina 224

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-24 2.4 ECOWATCH FOR WINDOWS -GETTING STARTED This sectio

Pagina 225

Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-3 INGREDIENTS: θ Potassium Chloride θ

Pagina 226

Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-4 INGREDIENTS: θ Ammonium Chloride θ

Pagina 227 - 4.3.5 REAL-TIME

Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-5 INGREDIENTS θ Potassium Nitrate θ

Pagina 228 - 4.3.6 GRAPH

Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-6 INGREDIENTS θ Styrene divinylbenzene c

Pagina 229 - 4.3.7 SETUP

Required Notice Appendix B YSI Incorporated Environmental Monitoring Systems Operations Manual B-1 APPENDIX B REQUIRED NOTICE The Federal Com

Pagina 230

Required Notice Appendix B YSI Incorporated Environmental Monitoring Systems Operations Manual B-2

Pagina 231

Accessories and Calibration Standards Appendix C YSI Incorporated Environmental Monitoring Systems Operations Manual C-1 APPENDIX C ACCESSORIES

Pagina 232

Accessories and Calibration Standards Appendix C YSI Incorporated Environmental Monitoring Systems Operations Manual C-2 θ 6628 Wiper Kit for 66

Pagina 233

Accessories and Calibration Standards Appendix C YSI Incorporated Environmental Monitoring Systems Operations Manual C-3 CONDUCTIVITY REAGENTS

Pagina 234

Accessories and Calibration Standards Appendix C YSI Incorporated Environmental Monitoring Systems Operations Manual C-4 θ 650-04 – High memory

Pagina 235

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-25 2.5 SONDE SOFTWARE SETUP There are two sets of softw

Pagina 236 - 4.3.8 APPL(ICATIONS)

Solubility and Pressure/Altitude Tables Appendix D YSI Incorporated Environmental Monitoring Systems Operations Manual D-1 APPENDIX D SOLUBILITY

Pagina 237 - 4.3.9 WINDOW

Solubility and Pressure/Altitude Tables Appendix D YSI Incorporated Environmental Monitoring Systems Operations Manual D-2 τ

Pagina 238 - 4.3.10 HELP

Solubility and Pressure/Altitude Tables Appendix D YSI Incorporated Environmental Monitoring Systems Operations Manual D-3 Table 2: Calibration V

Pagina 239 - 5.1 CONDUCTIVITY

Solubility and Pressure/Altitude Tables Appendix D YSI Incorporated Environmental Monitoring Systems Operations Manual D-4 Table 3. Conversion Fa

Pagina 240 - 5.2 SALINITY

Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-1 APPENDIX E TURBIDITY MEASUREMENTS Th

Pagina 241

Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-2 not recommend that you use tap water as

Pagina 242 - 5.5 pH

Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-3 For most applications involving both s

Pagina 243

Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-4 How do I manually activate the turbidit

Pagina 244 - 5.6 DEPTH AND LEVEL

Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-5 2. The probe is experiencing “interfer

Pagina 245 - POLAROGRAPHIC

Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-6

Pagina 246 -  Winkler titration

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-26  Set the Comm port number to match the port to which

Pagina 247 - FLOW DEPENDENCE

Flow Appendix F YSI Incorporated Environmental Monitoring Systems Operations Manual F-1 APPENDIX F FLOW Flow is only available in the Sensors

Pagina 248

Flow Appendix F YSI Incorporated Environmental Monitoring Systems Operations Manual F-2 Method Type Size Weir V-Notch 22½°, 30°, 45°, 60°, 90°,

Pagina 249

Flow Appendix F YSI Incorporated Environmental Monitoring Systems Operations Manual F-3 The final step is to return to the Report menu and choose

Pagina 250 - 5.10 NITRATE

Flow Appendix F YSI Incorporated Environmental Monitoring Systems Operations Manual F-4 Deploy the sonde so that the pressure sensor on the side

Pagina 251

Using Vented Level Appendix G YSI Incorporated Environmental Monitoring Systems Operations Manual G-1 APPENDIX G USING VENTED LEVEL Sondes that

Pagina 252 - 5.11 AMMONIUM AND AMMONIA

Using Vented Level Appendix G YSI Incorporated Environmental Monitoring Systems Operations Manual G-2 1. Place the short length of 1/4” tubing

Pagina 253

EMC Performance Appendix H YSI Incorporated Environmental Monitoring Systems Operations Manual H-1 APPENDIX H EMC PERFORMANCE When using the YS

Pagina 254 - 5.12 CHLORIDE

EMC Performance Appendix H YSI Incorporated Environmental Monitoring Systems Operations Manual H-2

Pagina 255 - 5.13 TURBIDITY

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-1 APPENDIX I CHLOROPHYLL MEASUREMEN

Pagina 256 - Near Infrared LED

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-2 6025 should approximately track the c

Pagina 257

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-27 Figure 39 If your sonde has previously been used, the

Pagina 258

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-3 Note: Be sure to carr

Pagina 259 - 5.14 CHLOROPHYLL

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-4 Will the chlorophyll values determine

Pagina 260 - IN VIVO

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-5 What Data Filter settings should I u

Pagina 261

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-6 available water sampling devices that

Pagina 262

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-7 Note also that your may observe less

Pagina 263

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-8 Note the significant improvement i

Pagina 264 - IN VIVO CHLOROPHYLL

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-9 Do I have to worry about the effect o

Pagina 265

Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-10

Pagina 266 - 5.15 RHODAMINE WT

Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J-1 APPENDIX J PERCENT AIR SATURATION The

Pagina 267 - Optical Filter

Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J-2 The “DOsat %Local” Convention In the s

Pagina 268 - PREPARATION

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-28 Figure 40 - Sonde Menu Flow Chart SYSTEM SETUP A

Pagina 269

Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J-3 %Local” values. For example, assume th

Pagina 270 - (Orange LED)

Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J-4

Pagina 271 - SENSOR DRIFT

PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K-1 APPENDIX K PAR SENSOR Some users of the YSI 6600-

Pagina 272

PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K-2 To set up your special PAR 66

Pagina 273

PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K-3 Finally, if you want to present your PAR data in ph

Pagina 274

PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K-4

Pagina 275 - GREEN ALGAE

Protective Zinc Anode Appendix L YSI Incorporated Environmental Monitoring Systems Operations Manual L-1 APPENDIX L PROTECTIVE ZINC ANODE Shipb

Pagina 276 - (Green LED)

Protective Zinc Anode Appendix L YSI Incorporated Environmental Monitoring Systems Operations Manual L-2 1. Using a large regular blade screwdr

Pagina 277

Protective Zinc Anode Appendix L YSI Incorporated Environmental Monitoring Systems Operations Manual L-3

Pagina 278

Protective Zinc Anode Appendix L YSI Incorporated Environmental Monitoring Systems Operations Manual L-4 NOTE: The zinc anode collar must be remo

Pagina 279

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-29 Select 1-Date & time. An asterisk will appear next

Pagina 280

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-1 APPENDIX M ROX OPTICAL DO SENSOR This

Pagina 281 - S•R•A•K=Q

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-2 3. The ROX sensor is automatically wipe

Pagina 282 - H•K+H•K=Q

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-3 • The 6150 optical sensor has no flow d

Pagina 283 - 6.1 CALIBRATION ERRORS

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-4 Should I calibrate my ROX sensor in wate

Pagina 284 - Troubleshooting Section 6

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-5 • When prompted for the input of a baro

Pagina 285

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-6 How often should I change my ROX sensor

Pagina 286

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-7 How should I store my ROX sensor when it

Pagina 287 - 7.2 HARDWARE INTERFACE

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-8 guard. It is also important to note th

Pagina 288 - 7.4 SDI-12 INTERFACE

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-9 protocol will prevent the stiff EDS auxi

Pagina 289

ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-10

Pagina 290 - Communications Section 7

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-30 ENABLING SENSORS To activate the sensors that are in y

Pagina 291

NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N-1 APPENDIX N NMEA APPLICATIONS 6-series so

Pagina 292

NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N-2 7, SpCond uS/cm 4, Cond mS/cm 5, Cond

Pagina 293 - 9.1 LIMITATION OF WARRANTY

NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N-3 204, Rhodamine ug/L 211, ODOsat % 214, ODOsa

Pagina 294 - 9.3 CLEANING INSTRUCTIONS

NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N-4

Pagina 295 - 9.4 PACKING INSTRUCTIONS

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-1 APPENDIX O SPECIFICATIONS SONDE SPECIFICATI

Pagina 296 - 9.5 PRODUCT RETURN FORM

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-2 Temperature: -5 to +50 C for most sensors Dep

Pagina 297

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-3 Computer Interface: RS-232C, SDI-12 Internal

Pagina 298

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-4 6920V2-2 SONDE Available Sensors Temperatur

Pagina 299

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-5 Storage Temperature: -40 to +60 C for sonde

Pagina 300 - YSI Zobell Solution: 3682

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-6 600XLM SONDE Available Sensors Temperature, C

Pagina 301

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-31 Note that the exact appearance of this menu will vary

Pagina 302

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-7 Internal logging memory size: 384 kilobytes

Pagina 303

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-8 Diameter: 1.65 inches (4.2 cm) Length:

Pagina 304

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-9 600LS SONDE Available Sensors Temperature, Co

Pagina 305

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-10 Non-Vented Level - Shallow Sensor Type...

Pagina 306 - CALIBRATION STANDARDS

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-11 pH Sensor Type...Glass combination ele

Pagina 307

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-12 Rhodamine WT Sensor type... Optical,

Pagina 308

Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-13 Ammonia-Nitrogen Sensor Type...Calcula

Pagina 309

Item # 069300 Drawing # A69300 Revision H November 2011 1700/1725 Brannum Lane Yellow Springs, Ohio 45387 USA (800) 765-4974 (937) 767-7241 FAX: (

Pagina 310 - PRESSURE/ALTITUDE TABLES

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-32 CHECKING ADVANCED SETTINGS Select Advanced from the Ma

Pagina 311

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-33 If you have a depth sensor installed, you can maintai

Pagina 312

5.14 CHLOROPHYLL 5-21 5.15 RHODAMINE 5-28 5.16 PHYCOCYANIN-CONTAINING BLUE-GREEN ALGAE 5-31 5.17 PHYCOERYTH

Pagina 313

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-34 2.6 GETTING READY TO CALIBRATE 2.6.1 INTRODUCTION

Pagina 314

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-35 6. Have several clean, absorbent paper towels or co

Pagina 315

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-36 RECOMMENDED VOLUMES OF CALIBRATION REAGENTS The approx

Pagina 316

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-37 Table 6 6600V2-4 Sonde with Standard Long Calibrati

Pagina 317

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-38 NOTE CAREFULLY: All optical sensors MUST be calibrate

Pagina 318

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-39 3. Note that the exact appearance of this menu will va

Pagina 319

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-40 Note that all parameters that have been enabled will

Pagina 320 - APPENDIX F FLOW

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-41 NOTE: The YSI conductivity system is very linear over i

Pagina 321 - MANNING EQUATION

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-42 Monitoring Applications If your instrument will be use

Pagina 322 - DEPLOYING YOUR SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-43 DEPTH AND LEVEL For the depth and level calibration, m

Pagina 323

Introduction Section 1 YSI Incorporated Environmental Monitoring Systems Manual 1-1 SECTION 1 INTRODUCTION 1.1 ABOUT YSI INCORPORATED From a

Pagina 324 - INSTALLING THE CARTRIDGE KIT

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-44 The next calibration instructions are only for the ISE

Pagina 325 - INSTALLING THE CANISTER KIT

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-45 CALIBRATION TIP: Exposure to the high ionic content

Pagina 326 - APPENDIX H EMC PERFORMANCE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-46 water. Input the value 0 NTU at the prompt, and press

Pagina 327 - EMC Performance Appendix H

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-47 Before making any field readings, carefully read Sectio

Pagina 328

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-48 RHODAMINE WT 2-POINT Select Rhodamine from the Calibra

Pagina 329

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-49 2.7 TAKING READINGS After you have (1) enabled the se

Pagina 330

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-50 If you started sampling without entering a filename, th

Pagina 331

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-51 sampling at 6:00 PM on July 31, 1996. The site is Clea

Pagina 332

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-52  Skip B-View params to log in this initial test stud

Pagina 333

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-53 The display now shows the next date and time for loggi

Pagina 334

Introduction Section 1 YSI Incorporated Environmental Monitoring Systems Manual 1-2 techniques, and system setup is necessary to obtain accurate

Pagina 335

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-54 2.8 USING ECOWATCH TO CAPTURE, UPLOAD AND ANALYZE DAT

Pagina 336

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-55 Select 1-Directory to view all fil

Pagina 337

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-56 Select 1-Proceed. Choose the appropriate file transfe

Pagina 338 - The “DOsat %” Convention

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-57 Select 5-Quick view file to view the last page of data

Pagina 339

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-58 Figure 41 Opening a File Some daily variations may b

Pagina 340

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-59 Figure 42 Viewing Options It may be somewhat awkward

Pagina 341

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-60 Figure 43 Viewing the Data in Table Format Viewing

Pagina 342 - APPENDIX K PAR SENSOR

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-61 Figure 44 Statistics and Study Information Next,

Pagina 343 - PAR Sensor Appendix K

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-62 Figure 45 Viewing the Data with Right-Button Mouse Fun

Pagina 344

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-63 Figure 46 Changing the Appearance of a Graph or Tabl

Pagina 345

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-1 SECTION 2 SONDES 2.1 GETTING STARTED The 6-Series En

Pagina 346

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-64 along the x-axis, you may also manually scale the y-axi

Pagina 347

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-65 Figure 48 Using Limit Data Set to Display a Subset of

Pagina 348

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-66 Figure 49 Saving, Exporting, Printing and Related Func

Pagina 349

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-67 Figure 50 Customizing a Graph from SAMPLE.DAT Final

Pagina 350

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-68 2.9 SONDE MENU The functions of the sondes a

Pagina 351

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-69 2.9.1 RUN Select 1-Run from the Main menu to begin t

Pagina 352

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-70 Select 1 – Start sampling option to start discrete samp

Pagina 353

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-71 Sampling Faster Than 0.5 Seconds For special applicati

Pagina 354

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-72 Follow the prompts on this screen to prepare your sond

Pagina 355

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-73  Select B-View Parameters to log to confirm that your

Pagina 356

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-2 2.2 CONNECTING YOUR SONDE There are a number of ways i

Pagina 357

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-74 Select 1-Yes and the screen will change. The displa

Pagina 358

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-75 2.9.2 CALIBRATE All of the sonde sensors (except temp

Pagina 359

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-76 CONDUCTIVITY Select 1-Conductivity to calibrate the c

Pagina 360

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-77 For the mg/L mode, calibration is carried out in a wate

Pagina 361

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-78 membrane installed on the 6150 probe and NOT a function

Pagina 362

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-79 which characterize the new membrane and they MUST be en

Pagina 363

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-80 Calibration Using mg/L – 1-Point Place the sensor in a

Pagina 364 - APPENDIX O SPECIFICATIONS

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-81 NOTE CAREFULLY: If you used sodium sulfite solution as

Pagina 365 - 6600V2-4 SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-82 pH When selecting ISE1 pH, you will be given the choic

Pagina 366 - 6920V2-1 SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-83 When selecting ISE3-NH4+, you will be given the choice

Pagina 367 - 6820V2-1 SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-3 Figure 2 Sonde to Data Collection PlatformYou will need

Pagina 368 - 6820V2-2 SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-84 CHLORIDE When selecting ISE5-CL-, you will be given th

Pagina 369 - 600XL SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-85 When selecting Optic X -6026-Turbidity (or 6136-Turbidi

Pagina 370 - 600QS SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-86 Enter after the readings have stabilized to confirm the

Pagina 371 - 600 OMS V2-1 SONDE

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-87 To calibrate your 6131 Blue-Green Algae-Phycocyanin (BG

Pagina 372 - Non-Vented Level – Deep

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-88 determining the BGA-PC content of algal suspensions by

Pagina 373

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-89 point calibration, but the software will prompt you to

Pagina 374 - in YSI AMCO-AEPA standards

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-90 CALIBRATION RECORD – THE GLP FILE When any sensor is c

Pagina 375

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-91 Pressing 1-View file will show the calibration record f

Pagina 376 - SOFTWARE SPECIFICATIONS

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-92 Constants and is described in Section 2.9.8. All other

Pagina 377 - Website: www.ysi.com

Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-93  PC6000 format will transfer the data so that it wil