Cond 3110 What to do if...
33
ba75792e02 11/2008
6 What to do if...
Error message
OFL, UFL
Error message,
E3
Sensor symbol flashes
Display,
LoBat
Meter does not react to
keystroke
You want to know which
software version is in
the meter
Cause Remedy
Measured value outside the
measuring range
Use suitable measuring cell
Cause Remedy
Measuring cell contaminated Clean cell and replace it if
necessary
Calibration solution not suitable Check calibration solutions
Cause Remedy
Cleaning interval expired Recalibrate the measuring
system
Cause Remedy
Batteries almost empty Replace the batteries (see
section 5.1 M
AINTENANCE)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press the <ENTER> and
<On/Off>
key
simultaneously
Cause Remedy
E. g., a question by the service
department
Switch on the meter.
During the display test,
display the software version
with <ENTER>.