Ion concentration ProLab 4000
130
ba75646e02 01/2010
11 In the Standard register, enter the values for Standard volume
and Standard concentration if necessary.
12 Add the standard solution to the sample.
13 Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
14 Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.
15 Change the displayed unit of the measured value with [Unit] if
necessary.
16 If required, display the record of the current measurement with
[Record].
17 Start measuring further samples with [Continue].
Repeat the steps 5 - 15 for all samples.
or
Terminate the measuring method with [Cancel] and confirm the
safety query with [OK].
Measurement with the selected method is completed.