Global Water
800-876-1172 globalw.com
Other Calculations:
Salinity:
Salinity is a unitless number but is generally referred to in PSS (Practical Salinity
Scale or PSU (Practical Salinity Units). Salinity can be calculated from
conductivity by the equation:
Salinity = A+(B*R^0.5)+(C*R)+(D*R^1.5)+(E*R^2)+(F*R^2.5)
Where:
A = 0.0120
B = -0.2174
C = 25.3283
D = 13.7714
E = -6.4788
F = 2.5842
R = Measured conductivity in uS / 53087uS
Altitude:
Calculating the barometric pressure based on altitude is not exact. It can vary with
air temperature, temperature lapse rate, and other atmospheric conditions. The
following is a good approximation.
Elevation
Feet
Pressure
mbar
Elevation
Feet
Pressure
mbar
Elevation
Feet
Pressure
mbar
0
1013
944
4000
878
250
1004
936
4250
870
500
994
926
4500
862
750
985
918
4750
854
1000
977
910
5000
846
1250
969
902
5250
838
1500
960
894
5500
832
1750
952
886
5750
824
P
a = 1013 - (3.54733E-2*A) + (4.6068714E-7 * A^2)
Where:
P
a = Atmospheric Pressure in mbar
A = Altitude in feet