Xylem Lab 850, starting from serial number 09470315 Manual de usuario Pagina 45

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 73
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 44
Lab 850 Operation
45
ba75890e01 11/2009
4.7.2 Measurement settings
These settings apply to calibration and measurement (the default
setting is printed in bold).
Buffer set for pH
calibration
(SET)
Number of the buffer set for pH
calibration (SET)
1 ... 16
Display during calibration (BUFFER) pH (buffer nominal value),
U (electrode voltage)
Unit of the value for the slope (SLO) %, mV/pH
Unit of the value for the zero point
(ASY)
pH, mV
Resolution (rES) HI (0.001), Lo (0.01)
Temperature unit (UnIt) °C, °F
Calibration interval (Int.C) 0 ... 7 ... 999 d
1 Open the menu for measurement settings with <MODE_>.
Set 1 ... 16, the specified buffer set is displayed.
2 Using <><>, select a buffer set Set 1 ... 16.
3 Confirm with <OK>.
bUFF, the setting of the display during calibration is displayed.
1
AutoCal
SET
Vista de pagina 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 72 73

Comentarios a estos manuales

Sin comentarios