ba75890e01 11/2009Lab 850 Operating manualLaboratory pH meter8422996pH°CTPLab 850

Pagina 2 - © 2009, SI Analytics GmbH

Overview Lab 85010ba75890e01 11/20091.3 Socket fieldCAUTIONOnly connect sensors to the meter that cannot return any voltages or currents that are not

Pagina 3 - DÉCLARATION DE CONFORMITÉ

Lab 850 Safety11ba75890e01 11/20092 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation a

Pagina 4

Safety Lab 85012ba75890e01 11/20092.1 Authorized useThis meter is authorized exclusively for pH and ORP measurements in the laboratory. The technical

Pagina 5 - Lab 850 - Contents

Lab 850 Safety13ba75890e01 11/2009Safe operation If safe operation is no longer possible, the instrument must be taken out of service and secured agai

Pagina 6 - Contents Lab 850

Safety Lab 85014ba75890e01 11/2009

Pagina 7 - 1Overview

Lab 850 Commissioning15ba75890e01 11/20093 Commissioning3.1 Scope of deliveryz Lab 850 laboratory meter z Power packz 4 batteries 1.5 V Mignon type AA

Pagina 8 - 1.1 Keypad

Commissioning Lab 85016ba75890e01 11/2009CAUTIONMake sure that the poles of the batteries are the right way round.The ± signs on the batteries must co

Pagina 9 - 1.2 Display

Lab 850 Operation17ba75890e01 11/20094 Operation4.1 Switching on the meterNoteThe meter has an energy saving feature to avoid unnecessary battery depl

Pagina 10 - 1.3 Socket field

Operation Lab 85018ba75890e01 11/20094.2 General operating principlesThis section contains basic information on the operation of the Lab 850. 4.2.1 Op

Pagina 11 - 2 Safety

Lab 850 Operation19ba75890e01 11/20094.3 MeasuringPreparatory activities Perform the following preparatory activities when you want to measure:NoteInc

Pagina 12 - 2.1 Authorized use

Lab 8502ba75890e01 11/2009Accuracy when going topressThe use of advanced technology and the high quality standard of our instruments are the result of

Pagina 13 - Lab 850 Safety

Operation Lab 85020ba75890e01 11/20094.3.1 Measuring the pH valueStability control SC(drift control)The stability control function (drift control) che

Pagina 14 - Safety Lab 850

Lab 850 Operation21ba75890e01 11/20094.3.2 Measuring the ORPThe meter can, in conjunction with an ORP electrode, measure the ORP (mV) of a solution. N

Pagina 15 - 3 Commissioning

Operation Lab 85022ba75890e01 11/20094.4 CalibrationWhy calibrate? pH electrodes age. This changes the zero point (asymmetry) and slope of the pH elec

Pagina 16 - Commissioning Lab 850

Lab 850 Operation23ba75890e01 11/20098 Mettler Toledo TEC * 1.9954.0057.0029.20825 °C9 Fisher * 2.0074.0027.00410.00225 °C10 Fluka BS * 4.0066.9848.95

Pagina 17 - 4 Operation

Operation Lab 85024ba75890e01 11/2009NoteThe buffer set is selected in the menu for measurement settings (see page 45).A list of the stored buffer set

Pagina 18 - Operation Lab 850

Lab 850 Operation25ba75890e01 11/2009Displaying thecalibration dataYou can display the data of the last calibration (see page 36). Subsequently, you c

Pagina 19 - 4.3 Measuring

Operation Lab 85026ba75890e01 11/2009Preparatory activitiesCalError CalError< -30 or > 30... -62 or ... -50 Eliminate the error according to pag

Pagina 20

Lab 850 Operation27ba75890e01 11/20094.4.1 Calibration interval (Int.C)The calibration interval and calibration evaluation are indicated on the displa

Pagina 21 - Lab 850 Operation

Operation Lab 85028ba75890e01 11/20094.4.2 Automatic calibration (AutoCal)For this procedure, use one, two or three buffer solutions of the selected b

Pagina 22 - 4.4 Calibration

Lab 850 Operation29ba75890e01 11/2009NoteHere you can cancel the calibration procedure with <MODE>. This corresponds to a single-point calibrati

Pagina 23

Lab 8503ba75890e01 11/2009KONFORMITÄTSERKLÄRUNGDECLARATION OF CONFORMITYDÉCLARATION DE CONFORMITÉWir erklären in alleiniger Verantwortung, dass das Pr

Pagina 24

Operation Lab 85030ba75890e01 11/2009NoteHere you can cancel the calibration procedure with <MODE>. This corresponds to a two-point calibration.

Pagina 25 - _> key

Lab 850 Operation31ba75890e01 11/20094.5 SavingThe pH meter has an internal data memory. It can store up to 800 datasets. A complete dataset consists

Pagina 26

Operation Lab 85032ba75890e01 11/2009 Message StoFull This message appears when all of the 800 memory locations are occupied.When the memory is full,

Pagina 27

Lab 850 Operation33ba75890e01 11/20094.5.2 Automatic storage at intervals (AutoStore)In order to store measured values automatically at certain time i

Pagina 28

Operation Lab 85034ba75890e01 11/2009Switching off theautomatic storingfunctionYou can terminate the automatic storing function in the following ways:

Pagina 29

Lab 850 Operation35ba75890e01 11/20094.5.3 Downloading the data memoryYou can download stored measurement datasets:z to the displayz to the interface

Pagina 30 - <▲><▼>

Operation Lab 85036ba75890e01 11/20094.5.4 Erasing the data memoryYou can erase all stored measurement datasets.4.5.5 Downloading calibration dataYou

Pagina 31 - 4.5 Saving

Lab 850 Operation37ba75890e01 11/2009Download to display viacalibration menuWhile the calibration data is being displayed, you can:z press <OK>

Pagina 32

Operation Lab 85038ba75890e01 11/20094.6 Transmitting dataThe meter has a RS232 interface (serial port).Via the interface, you can transmit data to a

Pagina 33

Lab 850 Operation39ba75890e01 11/20094.6.1 Options for data transmissionVia the RS 232 interface, you can transmit data to a PC or an external printer

Pagina 34

Lab 8504ba75890e01 11/2009

Pagina 35

Operation Lab 85040ba75890e01 11/20094.6.2 Automatically downloading measurement datasets at intervalsIn order to automatically download to the interf

Pagina 36

Lab 850 Operation41ba75890e01 11/2009Socket assignment(RS232)4.6.6 Operation with MultiLab pilotWith the aid of the MultiLab pilot software, you can r

Pagina 37

Operation Lab 85042ba75890e01 11/20094.7 SettingsYou can adapt the meter to your individual requirements. The settings are done in the following menus

Pagina 38 - 4.6 Transmitting data

Lab 850 Operation43ba75890e01 11/20094.7.1 System settingsThe default setting is printed in bold. Baud rate (Baud) Switch-off interval(t.OFF)Baud ra

Pagina 39 - _>. Then you

Operation Lab 85044ba75890e01 11/2009 Date and time4 Set the switch-off interval with <▲><▼>.5 Confirm with <OK>.Day.Month, the sett

Pagina 40

Lab 850 Operation45ba75890e01 11/20094.7.2 Measurement settingsThese settings apply to calibration and measurement (the default setting is printed in

Pagina 41

Operation Lab 85046ba75890e01 11/2009Display duringcalibration(bUFF)Unit of the value for theslope (SLO)4 Using <▲><▼>, select the display

Pagina 42 - 4.7 Settings

Lab 850 Operation47ba75890e01 11/2009Unit of the value for thezero point (ASY)Resolution(rES)8 Using <▲><▼>, select the unit for the zero

Pagina 43

Operation Lab 85048ba75890e01 11/2009Temperature unit (Unit)Calibrationinterval(Int.C) 12 Using <▲><▼>, toggle between °C and °F.13 Confir

Pagina 44

Lab 850 Operation49ba75890e01 11/20094.7.3 Interval for automatic storing (AutoStore)After setting the interval for automatic storing the current meas

Pagina 45

Lab 850 Contents5ba75890e01 11/2009Lab 850 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71.1 Keypa

Pagina 46

Operation Lab 85050ba75890e01 11/20094.7.4 Interval for automatic data transmissionThe interval for automatic data transmission serves to transmit the

Pagina 47

Lab 850 Operation51ba75890e01 11/20094.8 Reset4.8.1 Resetting calibration valuesThis function resets the calibration values to the default condition.

Pagina 48 - toggle between °C and °F

Operation Lab 85052ba75890e01 11/20094.8.2 Resetting all meter settingsThis function resets to the default condition all meter settings including the

Pagina 49 - to select an interval

Lab 850 Maintenance, cleaning, disposal53ba75890e01 11/20095 Maintenance, cleaning, disposal5.1 MaintenanceThe only maintenance activity required is r

Pagina 50

Maintenance, cleaning, disposal Lab 85054ba75890e01 11/20095.2 CleaningOccasionally wipe the outside of the meter with a damp, lint-free cloth. Disinf

Pagina 51 - 4.8 Reset

Lab 850 What to do if...55ba75890e01 11/20096 What to do if...Error messageErr1Error messageErr2Error messageErr4Error messageCalErrorCause RemedypH e

Pagina 52

What to do if... Lab 85056ba75890e01 11/2009No stable measuredvalue CalClock flashesDisplay,LoBat– Incorrect buffer solutions – Change calibration pro

Pagina 53 - 5.1 Maintenance

Lab 850 What to do if...57ba75890e01 11/2009Obviously incorrectmeasured valuesInstrument does notreact to keystrokeYou want to know whichsoftwareversi

Pagina 54 - 5.4 Disposal

What to do if... Lab 85058ba75890e01 11/2009

Pagina 55 - 6 What to do if

Lab 850 Technical data59ba75890e01 11/2009Stand October 30, 2009Translation of the legally binding German version7 Technical data7.1 General dataDimen

Pagina 56 - What to do if... Lab 850

Contents Lab 8506ba75890e01 11/20094.7.2 Measurement settings . . . . . . . . . . . . . . . . . . . . 454.7.3 Interval for automatic storing (AutoSto

Pagina 57 - Lab 850 What to do if

Technical data Lab 85060ba75890e01 11/2009Stand October 30, 2009Translation of the legally binding German versionGuidelinesand norms usedEMC EC guidel

Pagina 58

Lab 850 Technical data61ba75890e01 11/2009Stand October 30, 2009Translation of the legally binding German version7.2 Measuring ranges, resolution, acc

Pagina 59 - 7 Technical data

Technical data Lab 85062ba75890e01 11/2009Stand October 30, 2009Translation of the legally binding German version

Pagina 60 - Technical data Lab 850

Lab 850 Lists63ba75890e01 11/20098ListsThis chapter provides additional information and orientation aids.Abbreviations The list of abbreviations expla

Pagina 61 - Lab 850 Technical data

Lists Lab 85064ba75890e01 11/2009Abbreviations°C Temperature unit °Celsius°F Temperature unit, °FahrenheitASY Zero point (asymmetry)AutoCal Automatic

Pagina 62

Lab 850 Lists65ba75890e01 11/2009GlossaryAsymmetry see zero pointResolution Smallest difference between two measured values that can be displayed by a

Pagina 63 - Lab 850 Lists

Lists Lab 85066ba75890e01 11/2009Potentiometry Name of a measuring technique. The signal (depending on the measured parameter) of the electrode is the

Pagina 64 - Lists Lab 850

Lab 850 Lists67ba75890e01 11/2009Trademarks usedTrademark Owner of the trademarkMerck Merck KGaAMettler Toledo Mettler ToledoFisher Fisher Scientific

Pagina 65

Lists Lab 85068ba75890e01 11/2009

Pagina 66

Lab 850 ListsIndex69ba75890e01 11/2009AAuthorized use . . . . . . . . . . . . . . . . . . . . . . 12AutoCal . . . . . . . . . . . . . . . . . . . .

Pagina 67

Lab 850 Overview7ba75890e01 11/20091OverviewThe compact Lab 850 precision pH meter enables you to perform pH measurements rapidly and reliably. The La

Pagina 68

Lists Lab 85070ba75890e01 11/2009

Pagina 69

Lab 85071ba75890e01 11/2009Appendix: Firmware updateGeneral information With the Update_Labxxx_MxxxP program and a PC you can update the firmware of t

Pagina 70

Lab 85072ba75890e01 11/2009

Pagina 71 - Appendix: Firmware update

SI Analytics GmbHPostfach 2443D-55014 MainzHattenbergstr. 10D-55122 MainzEmail: support@Telefon +49 (0) 61 31/66 5111Telefax +49 (0) 61 31/66 5001si-a

Pagina 72

Overview Lab 8508ba75890e01 11/20091.1 KeypadIn this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <OK>) in

Pagina 73

Lab 850 Overview9ba75890e01 11/20091.2 DisplayFunction displayindicators888881888°C°FU pH%LoBatAutoStoreSCTPCalErrorTimeDay.MonthYearBaudNo.IdentmV/pH