What to do if... Lab 850
56
ba75890e01 11/2009
No stable measured
value
CalClock flashes
Display,
LoBat
Incorrect buffer solutions Change calibration
procedure
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions
Cause Remedy
Cause Remedy
pH electrode:
Junction contaminated Clean junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if
necessary
Electrode + test sample:
Conductivity too low Use a suitable electrode
Temperature too high
Organic liquids
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Batteries almost empty Replace the batteries (see
page 53)