Xylem MULTI 3400i pH/Cond MULTI-PARAMETER METERS Manual de usuario Pagina 44

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 83
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 43
Operation pH/Cond 3400i
44
4.5 Calibration intervals (Int 3, Int 5)
For each measured variable, a time interval is stored. When it has ex-
pired, you will be reminded to calibrate. After a calibration interval has
expired, the probe symbol of the relevant measured variable flashes. It
is still possible to measure. By calibrating the relevant probe, the func-
tion is reset and the interval starts anew.
The following calibration intervals are set in the factory:
Setting the calibration
interval
You can change each of these intervals (1 ... 999 days):
Measured
parameter
Designation Default setting
pH/ORP Int 3 7 days (d 7)
Conductivity Int 5 180 days (d 180)
1 Switch off the measuring instrument.
2 Press the <M> key and hold it down.
3 Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.
4 Press the <RUN/ENTER> key repeatedly until Int 3 or Int 5 and
the required measured variable pH or appears. Example:
5 Set the required interval (in days) until the next calibration with
<> <>.
6 Confirm with <RUN/ENTER>.
7 Switch to the measuring mode with <M>.
O
8
1
°
C
TP
ARng
mg/l
LoBat
Baud
Day.Month No.
Ident
Year
Sal
%
mbar
S
RCL
AR
Store
Auto
°
F
1/
cm
/
K
%
TDS
Sal
mV/pH
S/
m
cm
cm
M
Lin
Tref25
Tref20
nLF
AutoCal DIN
AutoCal TEC
7
t
3
n
1
d
Time
Cal
pH
Vista de pagina 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 82 83

Comentarios a estos manuales

Sin comentarios