pH/Cond 3400i Operation
51
Sample printout:
Outputting the
calibration data
on the display
No. 1:
09.03.99 17:10
pH 10.01 25 °C
Tman AR
Ident : 1
No. 2:
09.03.99 17:11
pH 10.01 24.7 °C
Tauto AR
Ident : 1
No. 3:
09.03.99 17:12
305.2 mV
Tauto
Ident : 13
...
1 Press the <RCL> key repeatedly until CAL disp appears on the
display.
2 Press the <RUN/ENTER> key.
The data of the last calibration of all measured variables ap-
pears in the following sequence:
z pH: Slope SLO and asymmetry ASY
z Cond: Cell constant C
Information concerning the calibration procedure is output as
well.
Store
Auto
Ident
S
1
LoBat
Tref25
nLF
/
K
%
Sal
°
AR
TP
C
M
m
%
mg/l
cm
S/
cm
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
mbar
OUpH
mV/pH
AutoCal DIN
AutoCal TEC
Time
Baud
Day.Month No.
Year
d
P
C
A
L
I
S
RCL