Vista de pagina 49
Operation pH/Cond 3400i
50
Note
If you want to search for a certain element of the data record (e.g. date),
proceed as follows:
Outputting stored data
to the interface
Note
You can cancel the transmission with <M> or <RUN/ENTER>.
The transmitted data contains the entire contents of the storage in in-
crementing order of the location numbers.
1 Using <RUN/ENTER>, select the element (e.g. date).
2Press <> or <> repeatedly until the required element ap-
pears on the display.
After approx. 2 s the temperature of the displayed measured
value appears.
1 Press the <RCL> key repeatedly until Sto SEr appears on the
display.
2 Press the <RUN/ENTER> key.
The complete contents of the storage are transmitted to the in-
terface. During the data transmission the instrument incre-
ments the storage numbers. After the data transmission, the
instrument automatically switches to the last active measure-
ment mode.