Dissolved oxygen (D.O.) MultiLab 4010-2
48 ba76142e01 05/2013
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
7.1.2 Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.
7.2 Calibration
7.2.1 Why calibrate?
D.O. sensors age. This changes the zero point (asymmetry) and slope of the
D.O. sensor. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope.
7.2.2 When to calibrate?
When the calibration interval has expired
When your accuracy requirements are especially high
Routinely within the framework of the company quality assurance
7.2.3 Calibration procedures
The MultiLab 4010-2 provides 2 calibration procedures:
3. Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
Measured parameter Time interval Stability in the time
interval
D.O. concentration 20 seconds
Δ : better than 0.03 mg/l
D.O. saturation 20 seconds
Δ : better than 0.4 %
D.O. partial pressure 20 seconds
Δ : Better than 0.8 mbar
Temperature 15 seconds
Δ : Better than 0.5 °C