MultiLab 4010-2 Settings
ba76142e01 05/2013 69
Temp. comp. (TC) /
Method
nLF
lin
off
Procedure for temperature compen-
sation (see section 8.2 T
EMPERATURE
COMPENSATION, page 57).
This setting is only available for the
measured parameters, ϰ and ρ.
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear temperature
compensation.
This menu item is only available when
the linear temperature compensation
is set.
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters, ϰ and ρ.
Multiplier for TDS 0.40 ... 1.00 Factor for TDS value
Reset - Resets all sensor settings to the deliv-
ery condition (see section 9.6.1
R
ESETTING THE MEASUREMENT SET-
TINGS, page 73).
Menu item Possible
setting
Explanation