ba75691e01 01/2007Oxi 1970iOperating manualRUN / ENTERAUTO READCALSTOMRCLON / OFFOxi 1970img/lO71 7 692°CARngTPREL1Portable oxygen measuring instrume

Pagina 2 - © Weilheim 2007, WTW GmbH

Safety Oxi 1970i10ba75691e01 01/20072.2 General safety instructionsThis instrument is built and inspected according to the relevant guide-lines and no

Pagina 3 - Oxi 1970i - Contents

Oxi 1970i Commissioning11ba75691e01 01/20073 Commissioning3.1 Scope of deliveryz Portable measuring instrument, Oxi 1970i with integrated recharge-abl

Pagina 4 - Contents Oxi 1970i

Commissioning Oxi 1970i12ba75691e01 01/2007Connecting the plug-inpower supply unit3.3 Initial commissioningPerform the following activities:z For main

Pagina 5 - 1Overview

Oxi 1970i Commissioning13ba75691e01 01/20074 Set the date of the current day with <▲> <▼>.5 Confirm with <RUN/ENTER>.The date (month

Pagina 6 - 1.1 Display

Commissioning Oxi 1970i14ba75691e01 01/20073.4 Sensor quiverTo store the sensors during field operation and to keep the sensor ele-ment moist, the qui

Pagina 7 - 1.2 Keypad

Oxi 1970i Operation15ba75691e01 01/20074 Operation4.1 Switching on the measuring instrumentNoteThe measuring instrument has an energy saving feature t

Pagina 8 - 1.3 Socket field

Operation Oxi 1970i16ba75691e01 01/20074.2 Measuring4.2.1 General informationYou can measure the following variables:z D. O. concentrationz D. O. satu

Pagina 9 - 2 Safety

Oxi 1970i Operation17ba75691e01 01/20074.2.2 Measuring the D. O. concentrationNoteWhen measuring the concentration of test samples with a salt content

Pagina 10 - Safety Oxi 1970i

Operation Oxi 1970i18ba75691e01 01/20074.2.3 Measuring the D. O. saturationYou can measure the D. O. saturation as follows: 4.2.4 Automatic selection

Pagina 11 - 3 Commissioning

Oxi 1970i Operation19ba75691e01 01/20074.2.5 AutoRead AR (drift control) and hold functionThe AutoRead function (drift control) checks the stability o

Pagina 12 - 3.3 Initial commissioning

Oxi 1970i2ba75691e01 01/2007Accuracy whengoing to pressThe use of advanced technology and the high quality standard of our instruments are the result

Pagina 13 - Oxi 1970i Commissioning

Operation Oxi 1970i20ba75691e01 01/20074.2.6 Sensor with stirrer (StirrOx G)Connection Connect the 2-pole stirrer plug to the stirrer output of the me

Pagina 14 - 3.4 Sensor quiver

Oxi 1970i Operation21ba75691e01 01/20074.3 CalibrationWhy calibrate? D. O. sensors age. This changes the slope of the D. O. sensor. Calibra-tion deter

Pagina 15 - 4 Operation

Operation Oxi 1970i22ba75691e01 01/2007Printing thecalibration recordThe calibration protocol contains the calibration data of the current cal-ibratio

Pagina 16 - 4.2 Measuring

Oxi 1970i Operation23ba75691e01 01/20074.3.1 Calibrating with the air calibration vesselProceed as follows to calibrate the instrument:NoteThe sponge

Pagina 17 - Oxi 1970i Operation

Operation Oxi 1970i24ba75691e01 01/2007NoteIn chapter 6 WHAT TO DO IF... page 49, you will find the measures to take for error elimination. 7 Switch t

Pagina 18 - Operation Oxi 1970i

Oxi 1970i Operation25ba75691e01 01/20074.3.2 Calibrating via a comparison measurementProceed as follows to calibrate the instrument:1 Connect the D. O

Pagina 19

Operation Oxi 1970i26ba75691e01 01/2007NoteIf the correction factor deviates from 1.000, the display of the measured value units flashes.8 Switch to t

Pagina 20

Oxi 1970i Operation27ba75691e01 01/20074.3.3 Entering the salt content (salinity)A salt content correction is required in the oxygen concentration mea

Pagina 21 - 4.3 Calibration

Operation Oxi 1970i28ba75691e01 01/20074.4 SavingThe measuring instrument has an internal data memory. It can store up to 500 datasets. A complete dat

Pagina 22

Oxi 1970i Operation29ba75691e01 01/2007Message StoFullStoFullStoFullStoFull This message appears when all of the 500 storage locations are occu-pied.Y

Pagina 23

Oxi 1970i Contents3ba75691e01 01/2007Oxi 1970i - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 Di

Pagina 24 - In chapter 6 W

Operation Oxi 1970i30ba75691e01 01/20074.4.2 Saving automaticallyThe save interval (Int 1) determines the chronological interval between automatic sa

Pagina 25

Oxi 1970i Operation31ba75691e01 01/2007NoteThe AutoStore function is interrupted if you start other functions, e.g. output the data storage. After the

Pagina 26

Operation Oxi 1970i32ba75691e01 01/20074.4.3 Outputting the data storageYou can output the contents of the data storage:z Stored data on the displayz

Pagina 27

Oxi 1970i Operation33ba75691e01 01/2007NoteIf you want to search for a certain element (e.g. date), proceed as fol-lows:Outputting stored datato the i

Pagina 28 - 4.4 Saving

Operation Oxi 1970i34ba75691e01 01/2007Sample printout:Outputting thecalibration dataon the display Device No.: 99990000No. 1: 09.03.02 17:10

Pagina 29 - StoFullStoFull

Oxi 1970i Operation35ba75691e01 01/2007Outputting thecalibration protocolon the interfaceNoteYou will find a sample calibration protocol in PRINTING

Pagina 30

Operation Oxi 1970i36ba75691e01 01/20074.4.4 Clearing the memoryWith this function, you can delete the stored data records. 500 storage locations will

Pagina 31

Oxi 1970i Operation37ba75691e01 01/20074.5 Transmitting dataYou have the following possibilities of transmitting data:z One of the following options:–

Pagina 32

Operation Oxi 1970i38ba75691e01 01/2007Setting theData transmissionintervalThe default setting for the interval is OFF. To start the data transmission

Pagina 33

Oxi 1970i Operation39ba75691e01 01/20074.5.2 Recorder (analog output)You can transmit data to a recorder via the analog output. Connect the analog out

Pagina 34

Contents Oxi 1970i4ba75691e01 01/20074.6 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 414.7 Reset . . . . . . .

Pagina 35

Operation Oxi 1970i40ba75691e01 01/20074.5.3 PC/external printer (RS232 interface)Via the RS 232 interface, you can transmit the data to a PC or an ex

Pagina 36

Oxi 1970i Operation41ba75691e01 01/20074.6 ConfigurationYou can adapt the measuring instrument to your individual require-ments. To do this, the follo

Pagina 37 - 4.5 Transmitting data

Operation Oxi 1970i42ba75691e01 01/2007Displaying theair pressureCalibration intervalAutoRange automaticselection of themeasurement range 6 Confirm wi

Pagina 38

Oxi 1970i Operation43ba75691e01 01/2007Date and time11 Set the date of the current day with <▲> <▼>.12 Confirm with <RUN/ENTER>.The

Pagina 39

Operation Oxi 1970i44ba75691e01 01/20074.7 Reset You can reset (initialize) the measurement parameters and the config-uration parameters separately fr

Pagina 40 - 3 Ground

Oxi 1970i Operation45ba75691e01 01/2007Resetting themeasuring parametersResetting the configura-tion parameters1 Press the <RUN/ENTER> key and h

Pagina 41 - 4.6 Configuration

Operation Oxi 1970i46ba75691e01 01/2007

Pagina 42

Oxi 1970i Maintenance, cleaning, disposal47ba75691e01 01/20075 Maintenance, cleaning, disposal5.1 MaintenanceThe measuring instrument is maintenance-f

Pagina 43

Maintenance, cleaning, disposal Oxi 1970i48ba75691e01 01/2007Measuring instrument, Dispose of the measuring instrument without the rechargeable batter

Pagina 44 - 4.7 Reset

Oxi 1970i What to do if...49ba75691e01 01/20076 What to do if...Display LoBat Instrument does notreact to keystroke Error message 0FL0FL0FL0FLError me

Pagina 45

Oxi 1970i Overview5ba75691e01 01/20071OverviewThe portable Oxi 1970i measuring instrument enables you to carry out oxygen measurements rapidly and rel

Pagina 46

What to do if... Oxi 1970i50ba75691e01 01/2007Error message E7E7E7E7AR flashescontinuously Measured value too lowMeasured value toohighDisplay totot

Pagina 47 - Ni-Cd Cd

Oxi 1970i What to do if...51ba75691e01 01/2007Message Sto Ful lStoFullStoFullStoFull You want to know whichsoftware version is inthe instrument Cause

Pagina 48

What to do if... Oxi 1970i52ba75691e01 01/2007

Pagina 49 - 6 What to do if

Oxi 1970i Technical data53ba75691e01 01/20077 Technical data7.1 General dataTest certificates cETLus, CEDimensions approx. 90 x 200 x 190 mmWeight app

Pagina 50 - What to do if... Oxi 1970i

Technical data Oxi 1970i54ba75691e01 01/2007Serial interface Automatic switchover when a PC or a printer is connected via the cable, AK 340/B or AK 32

Pagina 51 - Oxi 1970i What to do if

Oxi 1970i Technical data55ba75691e01 01/2007Internal resistance < 5 Ohm (current limited to max. 0.2 mA out-put current)

Pagina 52

Technical data Oxi 1970i56ba75691e01 01/2007Guidelinesand norms used7.2 Measuring ranges, resolutions, accuraciesMeasuring ranges,resolutionsNote: The

Pagina 53 - 7 Technical data

Oxi 1970i Technical data57ba75691e01 01/2007Air pressure correc-tionAutomatic through integratedpressure sensor in the range of 500 ... 1100 mbar

Pagina 54 - Technical data Oxi 1970i

Technical data Oxi 1970i58ba75691e01 01/2007

Pagina 55 - Oxi 1970i Technical data

Oxi 1970i Lists598ListsThis chapter provides additional information and orientation aids.Abbreviations The list of abbreviations explains the indicato

Pagina 56

Overview Oxi 1970i6ba75691e01 01/2007NoteIf you need further information or application notes, you can obtain the following material from WTW:z Applic

Pagina 57 - Automatic through integrated

Lists Oxi 1970i60AbbreviationsAR AutoRead (drift control)ARng Automatic range switchingMeasuring instrument measures with highest reso-lution°C Temper

Pagina 58

Oxi 1970i Lists61GlossaryAdjusting To manipulate a measuring system so that the relevant value (e. g. the displayed value) differs as little as possib

Pagina 59 - Oxi 1970i Lists

Lists Oxi 1970i62Salinity The absolute salinity SA of seawater corresponds to the relationship of the mass of dissolved salts to the mass of the solut

Pagina 60 - Lists Oxi 1970i

Oxi 1970i Index63ba75691e01 01/2007IndexAAnalog output . . . . . . . . . . . . . . . . . . . 39Authorized use . . . . . . . . . . . . . . . . . . .

Pagina 61

Index Oxi 1970i64ba75691e01 01/2007

Pagina 62

Wissenschaftlich-Technische Werkstätten GmbHDr.-Karl-Slevogt-Straße 1D-82362 WeilheimGermanyTel: +49 (0) 881 183-0+49 (0) 881 183-100Fax: +49 (0) 881

Pagina 63 - Oxi 1970i Index

Oxi 1970i Overview7ba75691e01 01/20071.2 KeypadKey functionsSwitch measuring instrument on/off <ON/OFF> Select the measuring mode <M>:– D.

Pagina 64 - Index Oxi 1970i

Overview Oxi 1970i8ba75691e01 01/20071.3 Socket fieldSensors You can use the following sensors with the Oxi 1970i:z Oxygen depth armaturez D. O. senso

Pagina 65

Oxi 1970i Safety9ba75691e01 01/20072 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation