Oxi 1970i Operation
23
ba75691e01 01/2007
4.3.1 Calibrating with the air calibration vessel
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet). Follow
the instructions in the OxiCal
®
-SL or DurOx 325 operating manual.
1 Connect the D. O. sensor to the measuring instrument.
2 Keep an air calibration vessel ready for use.
3 Put the D. O. sensor into the air calibration vessel.
4 Press the <CAL> key repeatedly until O
2
CAL appears.
5 Press the <RUN/ENTER> key. AutoRead is active, AR flashes.
6 As soon as a stable value is achieved, AR stops flashing. Cal-
ibration is finished; the sensor symbol indicates the relative
slope determined and the sensor evaluation (see page 22).