Oxi 1970i Operation
41
ba75691e01 01/2007
4.6 Configuration
You can adapt the measuring instrument to your individual require-
ments. To do this, the following parameters can be changed (the status
on delivery is marked in bold):
Note
You can leave the configuration menu at any time with <M> . The pa-
rameters that have already been changed are stored.
Baud rate
Baud rate 1200, 2400, 4800, 9600
Air pressure display Current value in mbar
(no input possible)
Calibration interval (Int 3) 1 ... 14 ... 999 d
AutoRange ARng On or off
Date/time Any
1 Switch off the measuring instrument.
2 Press the <M> key and hold it down.
3 Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.
4 Select the required baud rate with <> <>.
5 Confirm with <RUN/ENTER>. On the display, the current air
pressure in mbar appears.
REL1
Store
RCL
Sal
mg/l
ARng
O
TP
Oxi
Cal
AR
°
C
S
Year
Ident
No.
Day.Month
Tref25 Tref20
LoBat
Lin
nLF
Sal
pH
TEC
Auto
Time
cm
1/
K
/
%
REL2
pH/mV
m
cm
S/
%
8 0 04
Baud
Vista de pagina 40