Xylem WQ730 TURBIDITY SENSOR Manual de usuario Pagina 1

Busca en linea o descarga Manual de usuario para Herramientas Xylem WQ730 TURBIDITY SENSOR. Xylem WQ730 TURBIDITY SENSOR User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 12
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Global Water
800-876-1172 globalw.com
- 1 -
Global Water
Instrumentation, Inc.
11390 Amalgam Way
Gold River, CA 95670
T: 800-876-1172
Int’l: (916) 638-3429, F: (916) 638-3270
Turbidity Sensor: WQ730
02-002
11/30/09
Vista de pagina 0
1 2 3 4 5 6 ... 11 12

Indice de contenidos

Pagina 1 - Global Water

Global Water 800-876-1172 • globalw.com - 1 - Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-117

Pagina 2 - Table of Contents

Global Water 800-876-1172 • globalw.com - 10 - Step 7) Remove the sensor and rinse in distilled water. Dry the sensor and clean the lenses with

Pagina 3

Global Water 800-876-1172 • globalw.com - 11 - Appendix C: Sensor Measurement Diagram

Pagina 4

Global Water 800-876-1172 • globalw.com - 12 - Declaration of CE Conformity Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA

Pagina 5

Global Water 800-876-1172 • globalw.com - 2 - Congratulations on your purchase of the Global Water WQ730 Turbidity Sensor. This instrument has bee

Pagina 6

Global Water 800-876-1172 • globalw.com - 3 - I. Use and Maintenance a. The WQ730 Turbidity Sensor is a sensitive optical device and should be

Pagina 7

Global Water 800-876-1172 • globalw.com - 4 - II. Sensor Specifications a. Specifications. Outputs: Two, 4-20mA Ranges: 0-50 NTU

Pagina 8

Global Water 800-876-1172 • globalw.com - 5 - III. Trouble Shooting There are no user serviceable parts inside this sensor Issue: Sensor readi

Pagina 9 - (50) = _______

Global Water 800-876-1172 • globalw.com - 6 - IV. Warranty a. Global Water Instrumentation, Inc. warrants that its products are free from de

Pagina 10

Global Water 800-876-1172 • globalw.com - 7 - V. Appendix A: Turbidity Sensor Calibration Note: The following steps show how to calibrate the

Pagina 11

Global Water 800-876-1172 • globalw.com - 8 - Preparation: a. Set up a test stand with a rod and clamp to hold the sensor. Beakers will be plac

Pagina 12 - Declaration of CE Conformity

Global Water 800-876-1172 • globalw.com - 9 - Checking Calibration Step 1) Connect the sensor to the power supply and current meter in the follo