Xylem Oxi 3210 Dissolved oxygen (D.O.) meter Manual de usuario

Busca en linea o descarga Manual de usuario para Herramientas Xylem Oxi 3210 Dissolved oxygen (D.O.) meter. Xylem Oxi 3210 Dissolved oxygen (D.O.) meter User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir

Indice de contenidos

Pagina 1 - Oxi 3210

ba75795e02 03/2009Oxi 3210Operating manualDissolved oxygen (D.O.) meter15.11.200808:37MenuO27.9225.0mg/l[Sal]

Pagina 2 - © Weilheim 2009, WTW GmbH

Safety Oxi 321010ba75795e02 03/20092.1 Authorized useThe authorized use of the meter consists exclusively of the measurement of dissolved oxygen and t

Pagina 3 - Oxi 3210 - Contents

Oxi 3210 Safety11ba75795e02 03/2009Safe operation If safe operation is no longer possible, the meter must be taken out of service and secured against

Pagina 4 - Contents Oxi 3210

Safety Oxi 321012ba75795e02 03/2009

Pagina 5 - 1Overview

Oxi 3210 Commissioning13ba75795e02 03/20093 Commissioning3.1 Scope of deliveryz D.O. meter Oxi 3210 z 4 batteries 1.5 V Mignon type AAz Short instruct

Pagina 6 - 1.1 Keypad

Commissioning Oxi 321014ba75795e02 03/2009NoteAlternatively, you can also use Ni-MH rechargeable batteries (type Mignon AA). In order to charge the ba

Pagina 7 - 1.2 Display

Oxi 3210 Operation15ba75795e02 03/20094 Operation4.1 Switching on the meterSwitching on Press the <On/Off> key.The meter performs a self-test. T

Pagina 8 - 1.3 Socket field

Operation Oxi 321016ba75795e02 03/20094.2 General operating principlesThis section contains basic information on the operation of the Oxi 3210. Operat

Pagina 9 - 2 Safety

Oxi 3210 Operation17ba75795e02 03/20094.2.2 NavigationMeasured value display In the measured value display, you can z Open the menu for calibration an

Pagina 10 - ECHNICAL DATA

Operation Oxi 321018ba75795e02 03/2009z FunctionsFunctions are designated by the name of the function. They are immediately carried out by confirming

Pagina 11 - Oxi 3210 Safety

Oxi 3210 Operation19ba75795e02 03/20094.2.3 Example 1 on navigation: Setting the language1 Press the <On/Off> key.The measured value display app

Pagina 12 - Safety Oxi 3210

Oxi 3210ba75795e02 03/2009Accuracy whengoing to pressThe use of advanced technology and the high quality standard of our instruments are the result of

Pagina 13 - 3 Commissioning

Operation Oxi 321020ba75795e02 03/20095 Select the General submenu with <▲><▼>. The current selection is displayed with a frame.6 Open the

Pagina 14 - Commissioning Oxi 3210

Oxi 3210 Operation21ba75795e02 03/20099 Confirm the setting with <ENTER>. The meter switches to the measuring mode.The selected language is acti

Pagina 15 - 4 Operation

Operation Oxi 321022ba75795e02 03/20094.2.4 Example 2 on navigation: Setting the date and timeThe measuring instrument has a clock with a date functio

Pagina 16 - Operation Oxi 3210

Oxi 3210 Operation23ba75795e02 03/20094 Change and confirm the setting with <▲><▼> and <ENTER>.The minutes are highlighted.5 Change

Pagina 17 - Oxi 3210 Operation

Operation Oxi 321024ba75795e02 03/20094.3 Sensor-independent settingsThe Storage & config menu comprises the following settings:z System (see sect

Pagina 18

Oxi 3210 Operation25ba75795e02 03/20094.3.2 Data storageThis menu contains all functions to display, edit and erase stored measured values.NoteDetaile

Pagina 19 - Storage & config

Operation Oxi 321026ba75795e02 03/20094.4 Dissolved oxygen4.4.1 General informationYou can measure the following parameters:z D.O. concentration z D.O

Pagina 20

Oxi 3210 Operation27ba75795e02 03/20094.4.2 MeasuringYou can carry out D.O. measurements as follows:Selecting thedisplayedmeasured parameterYou can sw

Pagina 21

Operation Oxi 321028ba75795e02 03/2009Stability control(AutoRead )The stability control function (AutoRead) continually checks the stability of the me

Pagina 22 - 14:53:40

Oxi 3210 Operation29ba75795e02 03/20094.4.3 Settings for D.O. sensors (menu or measurement and calibration settings)Overview The following settings ar

Pagina 23

Oxi 3210 Contents3ba75795d02 03/2009Oxi 3210 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 Key

Pagina 24 - _>[Menu] key in

Operation Oxi 321030ba75795e02 03/2009Temperature unit °C °FTemperature unit, degrees Celsius or degrees Fahrenheit.All temperatures are displayed wit

Pagina 25

Oxi 3210 Operation31ba75795e02 03/20094.4.4 D.O. calibrationWhy calibrate? D.O. sensors age. This changes the slope of the D.O. sensor. Calibration de

Pagina 26 - 4.4 Dissolved oxygen

Operation Oxi 321032ba75795e02 03/2009Calibration evaluation After the calibration, the measuring instrument automatically evaluates the current statu

Pagina 27

Oxi 3210 Operation33ba75795e02 03/20094.4.5 Displaying calibration recordsDisplaying thecalibration recordThe calibration record of the last calibrati

Pagina 28 - Δ : Better than 1.2 mbar

Operation Oxi 321034ba75795e02 03/20094.5 Data storageYou can transmit measured values (datasets) to the data storage:z Manual storage (see section 4.

Pagina 29

Oxi 3210 Operation35ba75795e02 03/20094.5.1 Manual storageYou can transmit a measurement dataset to the data storage as follows.If the storage is full

Pagina 30

Operation Oxi 321036ba75795e02 03/20094.5.2 Displaying and editing the measurement data storageThe contents of the manual measurement data storage can

Pagina 31

Oxi 3210 Operation37ba75795e02 03/2009z Switch directly to the measured value display with <M>.z Quit the display and move to the next higher me

Pagina 32 - HAT TO DO IF

Operation Oxi 321038ba75795e02 03/20094.6.2 Resetting the system settingsThe following system settings can be reset to the delivery status:The system

Pagina 33 - _> key

Oxi 3210 Maintenance, cleaning, disposal39ba75795e02 03/20095 Maintenance, cleaning, disposal5.1 MaintenanceThe only maintenance activity required is

Pagina 34 - 4.5 Data storage

Contents Oxi 32104ba75795d02 03/20094.6 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 374.6.1 Resetting the me

Pagina 35

Maintenance, cleaning, disposal Oxi 321040ba75795e02 03/20095.2 CleaningOccasionally wipe the outside of the measuring instrument with a damp, lint-fr

Pagina 36 - _>[Menu] key in the

Oxi 3210 What to do if...41ba75795e02 03/20096 What to do if...Error message,OFLError message,ErrorSensor symbol flashesDisplayMeter does not react to

Pagina 37 - 4.6 Reset

What to do if... Oxi 321042ba75795e02 03/2009

Pagina 38 - _>[Menu]

Oxi 3210 Technical data43ba75795e02 03/20097 Technical data 7.1 General dataDimensions approx. 180 x 80 x 55 mmWeight approx. 0.4 kgMechanical structu

Pagina 39 - 5.1 Maintenance

Technical data Oxi 321044ba75795e02 03/20097.2 Measuring ranges, resolution, accuracyNoteThe accuracy values specified here apply exclusively to the m

Pagina 40 - 5.4 Disposal

Oxi 3210 Technical data45ba75795e02 03/2009FCC Class A Equipment StatementNote: This equipment has been tested and found to comply with the limits fo

Pagina 41 - AINTENANCE)

Technical data Oxi 321046ba75795e02 03/2009

Pagina 42 - What to do if... Oxi 3210

Oxi 3210 Lists47ba75795e02 03/20098Lists This chapter provides additional information and orientation aids.Specialist terms The glossary briefly expla

Pagina 43 - 7 Technical data

Lists Oxi 321048ba75795e02 03/2009Salinity The absolute salinity SA of seawater corresponds to the relationship of the mass of dissolved salts to the

Pagina 44 - Technical data Oxi 3210

Oxi 3210 ListsIndex49ba75795d02 03/2009AAir calibration beaker . . . . . . . . . . . . . . . . . 31Authorized use . . . . . . . . . . . . . . . . . .

Pagina 45 - Oxi 3210 Technical data

Oxi 3210 Overview5ba75795e02 03/20091OverviewThe Oxi 3210 compact precision dissolved oxygen (D.O.) meter enables you to perform D.O. measurements qui

Pagina 46

Lists Oxi 321050ba75795d02 03/2009

Pagina 48 - Lists Oxi 3210

Wissenschaftlich-Technische Werkstätten GmbHDr.-Karl-Slevogt-Straße 1D-82362 WeilheimGermanyTel: +49 (0) 881 183-0+49 (0) 881 183-100Fax: +49 (0) 881

Pagina 49

Overview Oxi 32106ba75795e02 03/20091.1 KeypadIn this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <ENTER>

Pagina 50

Oxi 3210 Overview7ba75795e02 03/20091.2 DisplayFunction displayindicators<ENTER>: <ENTER_>:Opens the menu for measurement settings / confi

Pagina 51

Overview Oxi 32108ba75795e02 03/20091.3 Socket fieldCautionOnly connect D.O. sensors to the meter that cannot return any voltages or currents that are

Pagina 52

Oxi 3210 Safety9ba75795e02 03/20092 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation a

Comentarios a estos manuales

Sin comentarios