Oxi 3210 What to do if...
41
ba75795e02 03/2009
6 What to do if...
Error message,
OFL
Error message,
Error
Sensor symbol flashes
Display
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause Remedy
Measured value outside the
measuring range
Use a suitable D.O. sensor
Cause Remedy
D.O. sensor contaminated Clean D.O. sensor and
replace it if necessary
Cause Remedy
Cleaning interval expired Recalibrate the measuring
system
Cause Remedy
Batteries almost empty Replace the batteries (see
section 5.1 M
AINTENANCE)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press the <ENTER> and
<On/Off> key
simultaneously
Cause Remedy
E. g., a question by the service
department
Switch on the meter.
Open the menu, <F1
_>/
[Menü] / Speicher & Konfig. /
System / Service
Information. The instrument
data is displayed.