Oxi 3210 Operation
29
ba75795e02 03/2009
4.4.3 Settings for D.O. sensors
(menu or measurement and calibration settings)
Overview The following settings are possible for D.O. sensors:
z Salinity correction
z Salinity (salinity equivalent)
z Calibration interval
z Automatic Stability control
Settings The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the <F1>/[menu] or <ENTER> key.
After completing the settings, switch to the measured value display with
<M>.
Menu item Possible
setting
Description
Calibration /
Calibration record
- Displays the calibration
record of the last calibration.
Calibration /
Calibration interval
1 ... 999 d Calibration interval for the
D.O. sensor (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.
Sal correction On
Off
Manual salt content
correction for concentration
measurements.
Salinity 0.0 ... 70.0 Salinity or salinity equivalent
for the salt content
correction.
This function is only
available for concentration
measurements if the
manual salt content
correction is switched on.
Stability control On /Off Switches on or off the
automatic stability control
during measurement (see
section 4.3.3)