Xylem ProLab 3000 Manual de usuario Pagina 104

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 202
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 103
Ion concentration ProLab 3000
104
ba75645e03 12/2012
9.3.1 Carrying out calibration
Use two to nine standard solutions in ascending or descending order.
Preparatory activities Perform the following preparatory activities when you want to calibrate:
Calibration
Note
When calibrating within the framework of a method measurement, it is
always a two-point calibration.
1 Connect an ion sensitive electrode to the meter.
2 Highlight the channel of the electrode in the measured value
display.
3 If necessary, select the ISE display with <MODE>.
4 If necessary, open the Measuring / ISE settings / Ion type menu
and select the ion type to be measured.
5 Keep the standard solutions ready.
6 If necessary, measure the temperature of the standard solu-
tions with a thermometer.
7 If necessary, change the unit of the measurement result and
calibration standards in the menu, Measuring/Units.
1 Perform the preparatory activities (see page 104).
2 Start the calibration with <CAL>.
The ISEX calibration window opens.
Vista de pagina 103
1 2 ... 99 100 101 102 103 104 105 106 107 108 109 ... 201 202

Comentarios a estos manuales

Sin comentarios