ProLab 3000 ORP voltage
87
ba75645e03 12/2012
8 ORP voltage
8.1 General information
You can measure the following variables:
ORP [mV]
Relative ORP [mV]
Temperature
measurement
The temperature value is measured and documented with all measure-
ments.
You have the following options to measure the temperature:
Automatic temperature measurement with
Temperature sensor (NTC 30 or Pt 1000) integrated in electrode
(if available).
the integrated temperature sensor of another electrode that is
connected and immersed in the sample at the same time.
an external temperature sensor (NTC 30 or Pt 1000).
Manual determination and input of the temperature.
The display of the temperature indicates the active temperature mea-
suring mode:
Temperature
sensor
Resolution of
the temp. dis-
play
Mode
yes 0.1 °C Automatic with temperature
sensor
- 1 °C Manual