ProLab 3000 pH
71
ba75645e03 12/2012
7.2.3 Measurement settings
All settings for pH measurements are in the Measuring menu.
Menu item Setting Description
Measuring
pH resolution High ()
Medium
Low
Resolution of the pH display:
High = 0.001
Medium = 0.01
Low = 0.1
Alternative TP () () = alternative temperature
sensor is used
() = no alternative tempera-
ture sensor is used
Setting of man. temp... -35 ... +150 °C Entry of the manually deter-
mined temperature. For mea-
surements without
temperature sensor only.
Reset pH Resets all sensor settings to
the delivery condition (see
page 65).
Recorder... Start recorder (see page 147).