Xylem MULTI 3400 MULTI-PARAMETER METERS Manual de usuario Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 85
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
Operation Multi 3400i
16
4.2 pH value / ORP voltage
4.2.1 General information
Preparatory activities Perform the following preparatory activities when you want to measure:
Note
Incorrect calibration of pH electrodes leads to incorrect measured val-
ues. Calibrate regularly before measuring. You can only connect elec-
trodes of the NTC30 type or without temperature sensor.
Warning
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
1 Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status dis-
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.
2 Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3 Calibrate or check the measuring instrument with the elec-
trode.
4Using < > <>, toggle between the pH or mV measuring
modes.
Vista de pagina 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 84 85

Comentarios a estos manuales

Sin comentarios