Xylem MULTI 3400 MULTI-PARAMETER METERS Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 85
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
Operation Multi 3400i
18
4.2.2 Measuring the pH value
AutoRead AR
(Drift control)
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.
For identical measurement conditions, the following criteria apply:
1 Perform the preparatory activities according to section 4.2.1.
2 Immerse the pH electrode in the test sample.
3 Press the <> <> keys until pH appears in the status display.
The pH value appears on the display.
4 When measuring without a connected temperature sensor:
Options:
z Determine the current temperature using a thermometer
and, while keeping the <RUN/ENTER> key depressed, en-
ter this temperature value with <> <>.
z TP display indicator not displayed, socket for the second
probe is free:
Connect the second probe (Oxi or Cond) and immerse it in
the same sample. TP flashes, the temperature is automati-
cally measured using the second probe.
Tref20
Lin
Cal
8
1
S
SalOU
%
mg/l
mV/pH
1/
cm
Sal
AutoCal DIN
Tref25
nLF
AutoCal TEC
RCL
Auto
Store
LoBat
ARng
AR
Time
Baud
Day.Month No.
Ident
Year
mbar
TDS
cm
M
°
F
/
K
%
S/
m
cm
8
4
2
9
9
6
°
C
TP
pH
Reproducibility Response time
Better than 0.02 > 30 seconds
1 Call up the pH measuring mode with <> <>.
2 Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
Vista de pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 84 85

Comentarios a estos manuales

Sin comentarios