Vista de pagina 72
Multi 3400i Technical data
73
Precision of
Cond (± 1 digit)
Cell constant,
calibrating
Cond reference
temperature
Serial interface Connection of the cable AK 340/B, AK 325/S
Nonlinear compensation
Accuracy Sample temperature
± 0.5 % 0 °C ... 35 °C
according to EN 27 888;
± 0.5 % 35 °C ... 50 °C
extended nLF function
acc. to
WTW measurements
SAL Range 0.0 ... 42.0
Accuracy Sample temperature
± 0.1 5 °C ... 25 °C
± 0.2 25 °C ... 30 °C
T [°C] ± 0.1
C [cm
-1
] 0.450 ... 0.500
Tref Can be set to 20 °C or 25 °C
Type RS232, data output
Baud rate Can be set to
1200, 2400, 4800, 9600 Baud
Data bits 8
Stop bits 2
Parity None
Handshake RTS/CTS + Xon/Xoff
Cable length Max. 15m