6.4 General errors
Display LoBat
Instrument does not
react
to keystroke
Display to
toto
to
Probe symbol flashes
Message Sto Ful l
StoFullStoFull
StoFull
Cause Remedy
– Batteries almost empty – Exchange the batteries (see
section 5.1 M
AINTENANCE)
Cause Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Hold the <CAL> key
depressed and switch the
instrument on
Cause Remedy
– Time-out of the interface – Check that the instrument is
connected
Cause Remedy
– Calibration interval expired – Recalibrate the measuring
system
Cause Remedy
– All memory locations are full – Output data storage and clear
data storage