
Multi 3400i Operation
29
4.3.4 AutoRead AR (Drift control)
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.
Criteria With identical measurement conditions, the following applies:
1 Call up the measuring mode with <M> and/or <▲> <▼> .
2 Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).
3 Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This mea-
sured value is transmitted to the interface.
4 If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5 To terminate the AutoRead function: Press the <AR> key.
8
LoBat
Time
Baud
Day.Month No.
Ident
Year
Sal
%
mbar
1
S
RCL
Store
Auto
Cal
°
F
1/
cm
/
K
%
TDS
UpH Sal
mV/pH
S/
m
cm
cm
M
Lin
Tref25
Tref20
nLF
AutoCal DIN
AutoCal TEC
6
7
1
0
8
O
°
C
mg/l
TP
ARng
AR
7
Measuring mode Reproducibility Response time
D. O. concentra-
tion
Better than 0.05 mg/l > 10 seconds
D. O. saturation in-
dex
Better than 0.6 % > 10 seconds
Comentarios a estos manuales